| Literature DB >> 15906979 |
Angela Elia1, Pietro Mario Lugarà, Corrado Giancaspro.
Abstract
A photoacoustic trace-gas sensor for the measurement of nitric oxide with a detection limit of 500 parts in 10(9) has been demonstrated. The radiation source was a thermoelectrically cooled distributed-feedback quantum-cascade laser operating in pulsed mode near 5.3 microm with an average laser power of 8 mW. A resonant photoacoustic cell was excited in its first longitudinal mode by the modulated laser light. Preliminary measurements have been performed to test the performance of our photoacoustic sensor; possible improvements to reach lower detection limits are discussed.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15906979 DOI: 10.1364/ol.30.000988
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776