Thermal comparison of the AdvanTec Legacy, Sovereign WhiteStar, and Millennium phacoemulsification systems.

Richard J Mackool1, Marc A Sirota.   

Abstract

PURPOSE: To determine the operating temperature of ultrasonic tips used with the Alcon AdvanTec Legacy, AMO Sovereign WhiteStar, and Bausch & Lomb Millennium phacoemulsification systems.
SETTING: Mackool Eye Institute, Astoria, New York, New York, USA.
METHODS: Thermal imaging of ultrasonic tips (Legacy, WhiteStar, and Millennium) was performed in air and in human cadaver eyes using a duty cycle of 33% (WhiteStar) and 50% (Millennium and Legacy). In vitro temperatures were measured with the tip centered in the incision and intentionally decentered against the side of the incision. The stroke length of each instrument was also measured, and the operating frequency of the Legacy was evaluated with the addition of a tangential load.
RESULTS: Open air and in vitro testing demonstrated that tip temperatures with the Legacy were consistently the lowest. Temperatures measured with the WhiteStar and Millennium systems were higher and generally similar to each other. At identical console power settings, the stroke length of the WhiteStar and Millennium tips was longer than that of the Legacy. The frequency of the Legacy handpiece did not change significantly (less than 200 Hz) under conditions of tangential tip loading.
CONCLUSIONS: At identical console power settings and similar console duty cycles, the temperature elevation of ultrasonic tips was least for the Legacy and greater for the WhiteStar and Millennium under all conditions. The causes of these findings appear to be the longer stroke length of the WhiteStar and Millennium and the underestimation of the duty cycle with the WhiteStar.

Entities:  

Mesh:

Year:  2005        PMID: 15899461     DOI: 10.1016/j.jcrs.2004.11.040

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Comparison of clear corneal incision injuries between torsional and conventional phacoemulsification.

Authors:  Xu Chen; Yinghong Ji; Yi Lu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-07-20       Impact factor: 3.117

2.  Heat profiling of phacoemulsification tip using a thermal scanning camera.

Authors:  Wei Kiong Ngo; Louis W Lim; Colin S H Tan; Wee Jin Heng
Journal:  Int Ophthalmol       Date:  2013-03-20       Impact factor: 2.031

3.  In water thermal imaging comparison of the Alcon legacy and AMO sovereign phacoemulsification systems.

Authors:  Kevin M Miller; Michael D Olson
Journal:  Open Ophthalmol J       Date:  2008-02-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.