Literature DB >> 15896

A transcutaneous PO2 electrode for continuous monitoring of fetal state during delivery.

L Stange, C Wussow.   

Abstract

A transcutaneous PO2 electrode has been designed and constructed for the continuous measurement of fetal oxygen tension during labor and delivery. The electrode, tested in vitro, has a low oxygen consumption, a short response time, 10 sec, and a sufficient stability, 5% over 3 hr. The in vivo performance of the electrode, tested in neonates, shows a good correlation between transcutaneous PO2 and PO2 measured in samples of arterial blood. The continous registration of fetal tcPO2 during labor and delivery is exemplified.

Entities:  

Mesh:

Substances:

Year:  1976        PMID: 15896     DOI: 10.1002/j.1879-3479.1976.tb00627.x

Source DB:  PubMed          Journal:  Int J Gynaecol Obstet        ISSN: 0020-7292            Impact factor:   3.561


  1 in total

1.  Adhesives for intact skin: a preliminary investigation.

Authors:  G S Margules; D L Harris
Journal:  Med Biol Eng Comput       Date:  1980-09       Impact factor: 2.602