| Literature DB >> 1589396 |
B S Kuo1, J C Poole, K K Hwang.
Abstract
A fully automated column-switching HPLC procedure has been developed and validated for quantitation of ML-1035 and its five metabolites in plasma employing direct injection. Plasma samples were injected onto a CN extraction column (4 x 4.6 mm, 5 microns) for micellar cleanup with 0.5% sodium dodecyl sulfate (SDS) in 50 mM phosphate. The proteinaceous components were solubilized and flushed out. The extracted compounds were then eluted by forward flush onto a C8 analytical column (150 x 4.6 mm, 5 microns) for further analysis using fluorescence detection (excitation, 308 nm; emission, 350 nm). After the subsequent washing and reequilibration with a sequence of three solvent mixtures, the extraction column was ready for the next injection. The limit of quantitation for all compounds of interest was about 10 to 15 ng/ml using 100 microliters of plasma. Excellent precision, accuracy, and linearity were obtained for all compounds over a range of 10 to 1500 ng/ml. The practicality of the HPLC method was also validated with plasma samples from dogs receiving ML-1035. Longevity for both extraction and analytical columns is excellent. Micellar cleanup coupled with the column-switching technique is a promising HPLC procedure when using direct injection of biological fluids.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1589396 DOI: 10.1023/a:1018948314823
Source DB: PubMed Journal: Pharm Res ISSN: 0724-8741 Impact factor: 4.200