| Literature DB >> 15891869 |
P Massaroti1, L A B Moraes, M A M Marchioretto, N M Cassiano, G Bernasconi, S A Calafatti, F A P Barros, E C Meurer, J Pedrazzoli.
Abstract
A liquid chromatographic-tandem mass spectrometric method (LC-MS/MS) for quantifying amlodipine in human plasma was developed and validated. Sample preparation was based on liquid-liquid extraction using NaOH and a mixture of ethyl acetate/hexane (80/20; v/v). Chromatography was performed on a C-18 analytical column and the retention times were 1.9 and 3.0 min for amlodipine and nimodipine (internal standard), respectively. The ionization was optimized using ESI(+) and enhanced selectivity was achieved using tandem mass spectrometric analysis via two MRM functions, 409 --> 238 and 418 --> 343 for amlodipine and nimodipine. The calibration curve ranged from 0.2 to 20.0 ng/mL. The inter-day precision and accuracy and the relative standard deviation (RSD) were <15%. The analyte was shown to be stable over the time-scale of the whole procedure. The robustness of the method was demonstrated by the good reproducibility of the results obtained during the analysis of clinical samples.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15891869 DOI: 10.1007/s00216-005-3227-z
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142