Literature DB >> 15891190

Aneurysmal pressure changes with nondetachable balloon placement and fluid infusion: rationale for liquid embolization.

Steven G Imbesi1, Kimberly Knox, Charles W Kerber.   

Abstract

BACKGROUND AND
PURPOSE: To improve the safety and efficacy of liquid embolization, we evaluated changes in pressures in experimental aneurysms.
METHODS: We created three replicas of a lateral sidewall aneurysm and placed them in a physiologic flow circuit. A 3 x 10-mm nondetachable balloon was positioned in the parent vessel across the aneurysmal neck. Intra-aneurysmal pressures were measured at baseline and after balloon inflation. Fluid was infused into the sac via a 1.45F microcatheter during inflation, and maximal pressures were noted. Measurements were repeated eight times in each aneurysm.
RESULTS: After balloon inflation, average intra-aneurysmal pressures increased: 12 mm Hg (13%, sigma(n - 1) = 0.46) for aneurysm 1 (baseline mean arterial pressure [MAP], 94 mm Hg), 15 mm Hg (58%, sigma(n - 1) = 0.88) for aneurysm 2 (baseline MAP, 26 mm Hg), and 15 mm Hg (58%, sigma(n - 1) = 0.92) for aneurysm 3 (baseline MAP, 26 mm Hg). During inflation and infusion, pressures increased slightly: 1.1 (0.94%, sigma(n - 1) = 0.64), 1.6 (3.9%, sigma(n - 1) = 1.1), and 1.9 (4.6%, sigma(n - 1) = 1.2) mm Hg for aneurysms 1, 2, and 3, respectively. Despite complete balloon occlusion of the distal aneurysmal neck, a channel between the proximal aneurysmal neck and the parent-vessel lumen persisted along the microcatheter. Fluid exited the sac via this channel, preventing a concomitant, significant increase in pressure during infusion.
CONCLUSION: Intra-aneurysmal pressure modestly increased with inflation of a parent-vessel balloon across the neck. When liquid was infused into the sac during inflation, further increases were minimal.

Entities:  

Mesh:

Year:  2005        PMID: 15891190      PMCID: PMC8158627     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  18 in total

1.  Carotid-cavernous fistula caused by a ruptured intracavernous aneurysm: endovascular treatment by electrothrombosis with detachable coils.

Authors:  G Guglielmi; F Viñuela; F Briganti; G Duckwiler
Journal:  Neurosurgery       Date:  1992-09       Impact factor: 4.654

2.  Utility of balloon-assisted Guglielmi detachable coiling in the treatment of 49 cerebral aneurysms: a retrospective, multicenter study.

Authors:  J P Cottier; A Pasco; S Gallas; J Gabrillargues; C Cognard; J Drouineau; L Brunereau; D Herbreteau
Journal:  AJNR Am J Neuroradiol       Date:  2001-02       Impact factor: 3.825

3.  Cyanoacrylate embolization of experimental aneurysms.

Authors:  Jean Raymond; France Berthelet; Anne-Cécile Desfaits; Igor Salazkin; Daniel Roy
Journal:  AJNR Am J Neuroradiol       Date:  2002-01       Impact factor: 3.825

4.  Endovascular treatment of experimental wide neck aneurysms: comparison of results using coils or cyanoacrylate with the assistance of an aneurysm neck bridge device.

Authors:  Jean Raymond; Igor Salazkin; Stavros Georganos; François Guilbert; Anne-Cécile Desfaits; Guylaine Gevry; Alain Weill; Daniel Roy
Journal:  AJNR Am J Neuroradiol       Date:  2002 Nov-Dec       Impact factor: 3.825

5.  n-Butyl 2-cyanoacrylate--substitute for IBCA in interventional neuroradiology: histopathologic and polymerization time studies.

Authors:  M F Brothers; J C Kaufmann; A J Fox; J P Deveikis
Journal:  AJNR Am J Neuroradiol       Date:  1989 Jul-Aug       Impact factor: 3.825

6.  Coiling of very large or giant cerebral aneurysms: long-term clinical and serial angiographic results.

Authors:  Menno Sluzewski; Tomas Menovsky; Willem Jan van Rooij; Douwe Wijnalda
Journal:  AJNR Am J Neuroradiol       Date:  2003-02       Impact factor: 3.825

7.  Endovascular treatment of posterior circulation aneurysms by electrothrombosis using electrically detachable coils.

Authors:  G Guglielmi; F Viñuela; G Duckwiler; J Dion; P Lylyk; A Berenstein; C Strother; V Graves; V Halbach; D Nichols
Journal:  J Neurosurg       Date:  1992-10       Impact factor: 5.115

8.  Technical feasibility of embolizing aneurysms with glue (N-butyl 2-cyanoacrylate): experimental study in rabbits.

Authors:  Dae Chul Suh; Kil Soo Kim; Soo Mee Lim; Hai Bin Shi; Choong Gon Choi; Ho Kyu Lee; Dong Man Seo
Journal:  AJNR Am J Neuroradiol       Date:  2003-09       Impact factor: 3.825

9.  Endovascular embolization of 150 basilar tip aneurysms with Guglielmi detachable coils: results of the Food and Drug Administration multicenter clinical trial.

Authors:  J M Eskridge; J K Song
Journal:  J Neurosurg       Date:  1998-07       Impact factor: 5.115

10.  Aneurysm flow dynamics: alterations of slipstream flow for neuroendovascular treatment with liquid embolic agents.

Authors:  Steven G Imbesi; Kimberly Knox; Charles W Kerber
Journal:  AJNR Am J Neuroradiol       Date:  2003 Nov-Dec       Impact factor: 3.825