Surface optimization technique for MammoSite breast brachytherapy applicator.

Michael Kirk1, Wen Chien Hsi, Adam Dickler, James Chu, Kambiz Dowlatshahi, Darius Francescatti, Cam Nguyen.   

Abstract

PURPOSE: We present a technique to optimize the dwell times and positions of a high-dose-rate (192)Ir source using the MammoSite breast brachytherapy applicator. The surface optimization method used multiple dwell positions and optimization points to conform the 100% isodose line to the surface of the planning target volume (PTV). METHODS AND MATERIALS: The study population consisted of 20 patients treated using the MammoSite device between October 2002 and February 2004. Treatment was delivered in 10 fractions of 3.4 Gy/fraction, twice daily, with a minimum of 6 h between fractions. The treatment of each patient was planned using three optimization techniques. The dosimetric characteristics of the single-point, six-point, and surface optimization techniques were compared.
RESULTS: The surface optimization technique increased the PTV coverage compared with the single- and six-point methods (mean percentage of PTV receiving 100% of the prescription dose was 94%, 85%, and 91%, respectively). The surface method, single-point, and six-point method had a mean dose homogeneity index of 0.62, 0.68, and 0.63 and a mean full width at half maximum value of 189, 190, and 192 cGy/fraction, respectively.
CONCLUSION: The surface technique provided greater coverage of the PTV than did the single- and six-point methods. Using the FWHM method, the surface, single-, and six-point techniques resulted in equivalent dose homogeneity.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15890576     DOI: 10.1016/j.ijrobp.2004.10.038

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Dosimetric impact of source-positioning uncertainty in high-dose-rate balloon brachytherapy of breast cancer.

Authors:  Yongbok Kim
Journal:  J Contemp Brachytherapy       Date:  2015-10-29

2.  WBRT vs. APBI: an interim report of patient satisfaction and outcomes.

Authors:  Samantha M Bitter; Patricia Heffron-Cartwright; Christopher Wennerstrom; Jared Weatherford; Douglas Einstein; Louis C Keiler
Journal:  J Contemp Brachytherapy       Date:  2016-02-09

3.  Evaluation of the interfractional biological effective dose (BED) variation in MammoSite high dose rate brachytherapy.

Authors:  Yongbok Kim; E Day Werts; Mark G Trombetta; Moyed Miften
Journal:  J Appl Clin Med Phys       Date:  2010-06-10       Impact factor: 2.102

4.  Comparison of single and multiple dwell position methods in MammoSite high dose rate (HDR) brachytherapy planning.

Authors:  Yongbok Kim; Mark G Trombetta; Moyed Miften
Journal:  J Appl Clin Med Phys       Date:  2010-05-28       Impact factor: 2.102

5.  Brachytherapy in accelerated partial breast irradiation (APBI) - review of treatment methods.

Authors:  Janusz Skowronek; Magdalena Wawrzyniak-Hojczyk; Kinga Ambrochowicz
Journal:  J Contemp Brachytherapy       Date:  2012-09-29

6.  Dosimetric evaluation of multilumen intracavitary balloon applicator rotation in high-dose-rate brachytherapy for breast cancer.

Authors:  Yongbok Kim; Mark G Trombetta
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.