Literature DB >> 15883065

Electromyographic investigation on handling forces of mechanically counterbalanced and sensor-servomotor-supported surgical microscopes.

Thomas Hoell1, Michaela Nagel, G Huschak, Andre Beier, Hans Joerg Meisel.   

Abstract

BACKGROUND: Recent developments in sensor-servomotor-driven microscopes again initiated a discussion on the value of such technology for practical use in neurosurgery. The need for less force in moving a motor-supported microscope is advantageous. However, questions remain if well-known difficulties in the past such as resonance phenomenon, loss of natural feeling, and unequal handling forces in different situations have been overcome by the new generation of sensor-servo-supported surgical microscopes.
METHODS: Handling forces of a mechanically counterbalanced neurosurgical microscope (Zeiss NC4, Zeiss, Oberkochen, Germany) were compared with those of a sensor-servo-supported neurosurgical microscope (Moeller HiR 20-1000, Moeller, Wedel, Germany). Handling forces were correlated with the surface electromyogram measurement of the muscle activity of 4 neurosurgeons. The activity of the forearm muscles was measured while handling the 2 different microscopes in standardized tests.
RESULTS: The electrophysiologic measurement revealed that significantly less muscle activity was required to handle the sensor-servo-driven microscope in all directions. The untrained surgeons profited less than the skilled ones. Differences were most evident with disbalanced microscopes.
CONCLUSIONS: With this technology, the neurosurgeons exerted less effort, especially in strenuous test situations where the single-handed use of the microscope was mandatory. The reduced muscle forces that move the sensor-servo-type microscope and the continuously balanced state that might help prevent unwanted correction movements will ease intraoperative handling in general.

Entities:  

Mesh:

Year:  2005        PMID: 15883065     DOI: 10.1016/j.surneu.2004.07.039

Source DB:  PubMed          Journal:  Surg Neurol        ISSN: 0090-3019


  1 in total

1.  Ergonomics-based Positioning of the Operating Handle of Surgical Microscopes.

Authors:  Satoru Shimizu; Hiroki Kuroda; Takahiro Mochizuki; Toshihiro Kumabe
Journal:  Neurol Med Chir (Tokyo)       Date:  2020-05-14       Impact factor: 1.742

  1 in total