Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

H Tsumura1, N Kaku, T Torisu.   

Abstract

PURPOSES: To examine radiographically the component motion in a bipolar prosthesis and to determine whether the self-centering mechanism really works in vivo.
METHODS: 38 patients with 41 bipolar hip endoprostheses (30 for coxarthrosis and 11 for osteonecrosis of femoral head) were included in this study. Two radiographs of each case were taken to evaluate the self-centering mechanism. The first anteroposterior radiograph of both hip joints was taken at the maximum abduction while the patient standing on the endoprosthetic leg. The second radiograph was taken after the patient returned to neutral position while standing on 2 legs. In the present study, the order in which the radiographs were taken differed from previously reported studies. The radiographs were analysed using the method similar to that of Drinker and Murray. The adductive motion from abduction to a neutral position is within the range of inner bearing oscillation.
RESULTS: The outer head alignment changed from 23 degrees to 12 degrees in the patients with osteonecrosis. However, the valgus position of the outer head (36 degrees) remained unchanged in the patients with coxarthrosis standing on 2 legs in the neutral position.
CONCLUSION: The self-centering mechanism of the bipolar endoprosthesis functioned in the patients with osteonecrosis, but did not work in the coxarthrosis group.

Entities:  

Mesh:

Year:  2005        PMID: 15872400     DOI: 10.1177/230949900501300108

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  2 in total

1.  Long-term clinical results of bipolar hemiarthroplasty for osteoarthritis and rheumatoid arthritis of the hip: A retrospective study.

Authors:  Tetsutaro Abe; Nobuhiro Kaku; Tomonori Tabata; Hiroaki Tagomori; Hiroshi Tsumura
Journal:  J Orthop       Date:  2019-08-12

2.  Radiographic evaluation of linear wear of bipolar hemiarthroplasty devices in vivo.

Authors:  Nobuhiro Kaku; Shouhei Noda; Tomonori Tabata; Hiroaki Tagomori; Hiroshi Tsumura
Journal:  J Clin Orthop Trauma       Date:  2018-07-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.