| Literature DB >> 1587080 |
Abstract
The 'sensitivity region' method of back projection is converted from polar drive to adjacent drive current injection strategy and individual sensitivity coefficients incorporated. For each projection sensitivity coefficients for each pixel were established and pixels were grouped together into the 'sensitivity regions'. These contain all pixels that have their largest sensitivity coefficients associated with a particular measurement pair. Back projection of measurement data was performed using these coefficients as the basis of the reconstruction and less sensitive measurement pairs were ignored. Formulae for the back projection are discussed and images from computer simulated data and raw data supplied by Sheffield as part of this concerted action are presented. Comparison with polar current injection strategy is discussed.Mesh:
Year: 1992 PMID: 1587080 DOI: 10.1088/0143-0815/13/a/020
Source DB: PubMed Journal: Clin Phys Physiol Meas ISSN: 0143-0815