Literature DB >> 15869671

New simplified technique for 3D mapping and ablation of right ventricular outflow tract tachycardia.

Moeen A Saleem1, Scott Burkett, Rod Passman, Samer Dibs, Erica D Engelstein, Alan H Kadish, Jeffrey J Goldberger.   

Abstract

OBJECTIVE: To evaluate the safety and efficacy of using a circular multielectrode catheter for mapping and ablation of ventricular tachycardia (VT) or premature ventricular complexes (PVCs) from the right ventricular outflow tract (RVOT).
BACKGROUND: Three-dimensional (3D) mapping systems are commonly used for mapping and ablation of RVOT VT and PVCs. Newer catheters that are circular with multiple electrodes, such as the Lasso catheter, are capable of simultaneously recording from multiple points within a circumferential plane. Given the tubular structure of the RVOT, these catheters could be used for mapping tachycardias from the RVOT.
METHODS: A retrospective cohort study of patients undergoing radiofrequency (RF) ablation of RVOT VT or PVCs was performed. In group 1 (n = 7), mapping was performed with a single ablation catheter and fluoroscopy. In group 2 (n = 10), 3D mapping using ESI (n = 9) or CARTO (n = 1) was performed. In group 3 (n = 12), mapping was performed with a circular multielectrode catheter (n = 12). All ablations were performed with 4-mm tip catheters using RF energy.
RESULTS: Catheter ablation for RVOT VT (n = 15) or PVCs (n = 14) was performed on 29 cases in 26 patients, 9 males. Mean age was 35.9 years. In groups 1, 2, and 3, the mean number of lesions was 17.7 +/- 7.7, 13.6 +/- 7.7, and 18.2 +/- 22.7 and the median number of lesions was 20, 13, and 5, respectively. There were no significant differences in the number of lesions, RF time, fluoroscopy time, procedure time, and acute success rate among the three techniques. There were three complications in group 2 and one in group 3.
CONCLUSION: The use of a circular multielectrode catheter is as effective as the other standard available 3D mapping techniques, both in terms of procedural success and procedural characteristics. Additionally, because of the lower cost associated with using the circular multielectrode catheter approach, further evaluation should be performed to determine whether this is the most cost-effective approach to 3D mapping and ablation of RVOT tachycardias.

Entities:  

Mesh:

Year:  2005        PMID: 15869671     DOI: 10.1111/j.1540-8159.2005.09547.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Tissue tracking imaging for identifying the origin of idiopathic ventricular arrhythmias: a new role of cardiac ultrasound in electrophysiology.

Authors:  Sunil Roy; Vikram Sankar; Johnson Francis; Hiroshi Tada
Journal:  Indian Pacing Electrophysiol J       Date:  2005-07-01

2.  Radiofrequency catheter ablation of idiopathic right ventricular outflow tract arrhythmias.

Authors:  Naiara Calvo; Monique Jongbloed; Katja Zeppenfeld
Journal:  Indian Pacing Electrophysiol J       Date:  2013-01-01

Review 3.  Mapping and Ablation of Idiopathic Ventricular Fibrillation.

Authors:  Ghassen Cheniti; Konstantinos Vlachos; Marianna Meo; Stephane Puyo; Nathaniel Thompson; Arnaud Denis; Josselin Duchateau; Masateru Takigawa; Claire Martin; Antonio Frontera; Takeshi Kitamura; Anna Lam; Felix Bourier; Nicolas Klotz; Nicolas Derval; Frederic Sacher; Pierre Jais; Remi Dubois; Meleze Hocini; Michel Haissaguerre
Journal:  Front Cardiovasc Med       Date:  2018-09-18
  3 in total