Literature DB >> 15866953

Device for zone-II flexor tendon repair. A multicenter, randomized, blinded, clinical trial.

Brian W Su1, Michael Solomons, Andrew Barrow, Matshediso E Senoge, Marco Gilberti, Lawrence Lubbers, Edward Diao, H Matthew Quitkin, Melvin P Rosenwasser.   

Abstract

BACKGROUND: The stainless-steel Teno Fix tendon-repair device has improved biomechanical characteristics compared with those of suture repair, and it was well tolerated in a canine model. The purpose of this study was to compare the Teno Fix with suture repair in a clinical setting.
METHODS: Sixty-seven patients with isolated zone-II flexor tendon injury were randomized to be treated with a Teno Fix or a four-stranded cruciate suture repair. There were eighty-five injured digits: thirty-four were treated with the Teno Fix, and fifty-one served as controls. A modified Kleinert rehabilitation technique was employed, with active flexion starting at four weeks postoperatively. Patients were followed for six months by blinded observers who determined the range of motion, Disabilities of the Arm, Shoulder and Hand (DASH) score, pinch and grip strength, and pain score on a verbal scale and assessed swelling and neurologic recovery. Adverse outcomes, including device migration and rupture, were monitored at frequent intervals.
RESULTS: Nine of the fifty-one suture repairs ruptured, whereas none of the Teno Fix repairs ruptured (p < 0.01). Five of the nine ruptures were caused by resistive motion against medical advice. There were no differences between the two groups in terms of range of motion, DASH score, pinch and grip strength, pain, swelling, or neurologic recovery. The Teno Fix group had slightly slower resolution of pain and swelling compared with the control group. Of the patients who were available for follow-up at six months, sixteen of the twenty-four treated with a Teno Fix repair and nineteen of the twenty-seven treated with a control repair had a good or excellent result. One Teno Fix device migrated and extruded secondary to a wound infection. Of all eighty-five digits that were operated on, four were thought to have tendons of inadequate size to accommodate the device and nine were deemed to have inadequate exposure to allow placement of the anchors.
CONCLUSIONS: The Teno Fix is safe and effective for flexor tendon repair if the tendon size and exposure are sufficient. Tendon repairs with the Teno Fix have lower rupture rates and similar functional outcomes when compared with conventional repair, particularly in patients who are noncompliant with the rehabilitation protocol.

Entities:  

Mesh:

Year:  2005        PMID: 15866953     DOI: 10.2106/JBJS.C.01483

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  12 in total

1.  The resurgence of barbed suture and connecting devices for use in flexor tendon tenorrhaphy.

Authors:  Yazeed Mazen Gussous; Chunfeng Zhao; Peter C Amadio; Kai-Nan An
Journal:  Hand (N Y)       Date:  2011-06-30

2.  The use of Teno Fix tendon repair device in a patient with multiple flexor tendon ruptures.

Authors:  Wade Daniel Kubat; John Hsu; Anahita Azharian; Mark Rekant
Journal:  J Hand Microsurg       Date:  2010-08-11

Review 3.  Suture techniques for tendon repair; a comparative review.

Authors:  Shelley Rawson; Sarah Cartmell; Jason Wong
Journal:  Muscles Ligaments Tendons J       Date:  2013-08-11

4.  Rehabilitation following surgery for flexor tendon injuries of the hand.

Authors:  Susan E Peters; Bhavana Jha; Mark Ross
Journal:  Cochrane Database Syst Rev       Date:  2021-01-13

5.  Tendon-holding capacities of two newly designed implants for tendon repair: an experimental study on the flexor digitorum profundus tendon of sheep.

Authors:  Ismail Ağır; Mahmut Nedim Aytekin; Onur Başçı; Barış Caypınar; Bülent Erol
Journal:  Open Orthop J       Date:  2014-06-27

6.  Sutured tendon repair; a multi-scale finite element model.

Authors:  Shelley D Rawson; Lee Margetts; Jason K F Wong; Sarah H Cartmell
Journal:  Biomech Model Mechanobiol       Date:  2014-05-20

7.  Flexor tendon injuries of the hand treated with TenoFix: mid-term results.

Authors:  Lorenzo Rocchi; Antonio Merolli; Andrea Genzini; Gianfranco Merendi; Francesco Catalano
Journal:  J Orthop Traumatol       Date:  2008-07-17

Review 8.  ICF components of corresponding outcome measures in flexor tendon rehabilitation - a systematic review.

Authors:  Renée Oltman; Gudrun Neises; Daniel Scheible; Gerhard Mehrtens; Christian Grüneberg
Journal:  BMC Musculoskelet Disord       Date:  2008-10-15       Impact factor: 2.362

Review 9.  The evidence-base for the management of flexor tendon injuries of the hand: Review.

Authors:  Athanasius Ishak; Akshaya Rajangam; Ankur Khajuria
Journal:  Ann Med Surg (Lond)       Date:  2019-10-15

10.  Functional and Disability Assessment Among Hispanics With Zone 2 Flexor Tendon Injuries: Comparative Study Between Flexor Digitorum Superficialis Repair and Flexor Digitorum Superficialis Excision.

Authors:  Eduardo J Natal-Albelo; Gerardo Olivella; Giovanni U Paraliticci-Márquez; Lenny Rivera; Gabriel Echegaray; Norman Ramírez; Christian A Foy-Parrilla
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2020-09