Comparison between flutter valve drainage bag and underwater seal device for pleural drainage after lung surgery.

S Vuorisalo1, P Aarnio, J Hannukainen.   

Abstract

AIM OF THE STUDY: The purpose of the present study was to compare the use of a conventional underwater seal device with suction and a flutter valve drainage bag for pleural drainage after lung surgery. PATIENTS AND METHODS: Altogether 59 patients undergoing elective lung surgery except pneumonectomy between February 2001 and April 2002 were prospectively randomized to receive postoperative pleural drainage by 28F chest tube(s) attached to underwater seal device placed on negative pressure of 15 cm of water or flutter valve drainage bag. Following withdrawal of four patients from the study, 55 patients were evaluated (31 patients in the underwater seal device group and 24 patients in the flutter valve drainage bag group).
RESULTS: In the conventional underwater seal device group the mean drainage time was 2.6 (SD +/- 2.0) days and in the flutter valve drainage bag group the mean drainage time was 3.3 days (SD +/- 4.0); difference -0.8, 95% confidence interval (CI) -2.4 to 0.9. The mean length of hospitalization in the surgical ward was 3.6 (SD +/- 2.7) and 4.1 (SD +/- 4.4) days respectively (difference -0.5, 95% CI -2.5 to 1.4).
CONCLUSIONS: The results of this study suggest that flutter valve drainage system is a safe and feasible alternative in managing postoperative air leaks and haemorrhage after lung surgery other than pneumonectomy if air leaks are not extremely massive.

Entities:  

Mesh:

Year:  2005        PMID: 15865119     DOI: 10.1177/145749690509400114

Source DB:  PubMed          Journal:  Scand J Surg        ISSN: 1457-4969            Impact factor:   2.360


  2 in total

1.  Promising Effects of Digital Chest Tube Drainage System for Pulmonary Resection: A Systematic Review and Network Meta-Analysis.

Authors:  Po-Chih Chang; Kai-Hua Chen; Hong-Jie Jhou; Cho-Hao Lee; Shah-Hwa Chou; Po-Huang Chen; Ting-Wei Chang
Journal:  J Pers Med       Date:  2022-03-22

2.  Comparison between flutter valve drainage bag and underwater seal device for the management of non-massive malignant and paramalignant pleural effusions.

Authors:  Olugbenga Olalekan Ojo; Martins Oluwafemi Thomas; Ezekiel Ogunleye; Olugbenga Olusoji; Uvie Ufuoma Onakpoya
Journal:  Pan Afr Med J       Date:  2020-01-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.