Use of solid-phase microextraction for the detection of acetic acid by ion-trap gas chromatography-mass spectrometry and application to indoor levels in museums.

Ana F L Godoi1, Luc Van Vaeck, René Van Grieken.   

Abstract

A simple and efficient method using solid-phase microextraction (SPME) and gas chromatography-ion trap mass spectrometry (GC-ITMS) was developed for the analysis of acetic acid in air. The choice of the SPME fibre revealed to be critical as well as the sampling and desorption time. A dilution vessel was used for calibration. The precision of the method was found to be 4.7% relative standard deviation (RSD) and the detection limit 5.7 microg m(-3). The SPME-GC-MS technique was applied to the analysis of acetic acid in museum atmospheres.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15844539     DOI: 10.1016/j.chroma.2004.12.054

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Validation Study of Selected Ion Flow Tube-Mass Spectrometry (SIFT-MS) in Heritage Science: Characterization of Natural and Synthetic Paint Varnishes by Portable Mass Spectrometry.

Authors:  Jacopo La Nasa; Francesca Modugno; Maria Perla Colombini; Ilaria Degano
Journal:  J Am Soc Mass Spectrom       Date:  2019-09-05       Impact factor: 3.109

2.  Non-Destructive Evaluation of Historical Paper Based on pH Estimation from VOC Emissions.

Authors:  Matija Strlič; Irena Kralj Cigić; Jana Kolar; Gerrit De Bruin; Boris Pihlar
Journal:  Sensors (Basel)       Date:  2007-12-05       Impact factor: 3.576

  2 in total