| Literature DB >> 15844324 |
Gossett A Campbell1, Raj Mutharasan.
Abstract
A composite self-excited millimeter-sized lead zirconate titanate (PZT) glass cantilever (2 mm x 1.8 mm; sensing area of 6 mm2) was fabricated for the detection of Escherichia coli (E. coli) O157:H7. The fundamental and second mode resonance in air was 10.95 +/- 0.05 kHz and 43.45 +/- 0.05 kHz, respectively. Affinity purified monoclonal antibody (anti-E. coli O157:H7) specific to the pathogen E. coli O157:H7 was immobilized at the cantilever glass tip, and then immersed in liquid containing the pathogen (70 to 7 x 10(7) cells/mL). The resonant frequency showed a reduction and reached a steady state shift of 0 +/- 5, 46 +/- 5, 260 +/- 5, and 1010 +/- 5 Hz corresponding to 0, 700, 7000, and 7 x 10(7) cells/mL. From the experiments conducted, the detection limit of the sensor was 700 cells/mL.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15844324 DOI: 10.2116/analsci.21.355
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081