Literature DB >> 15835112

Adjustment and malfunction of a programmable valve after exposure to toy magnets. Case report.

Richard C E Anderson1, Marion L Walker, John M Viner, John R W Kestle.   

Abstract

Inadvertent adjustments and malfunctions of programmable valves have been reported in cases in which patients have encountered powerful electromagnetic fields such as those involved in magnetic resonance imaging, but the effects of small magnetic fields are not well known. The authors present a case in which a child playing with a collection of commercially available toy magnets altered the pressure setting of an implanted valve and may have caused its permanent malfunction.

Entities:  

Mesh:

Year:  2004        PMID: 15835112     DOI: 10.3171/ped.2004.101.2.0222

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  11 in total

1.  The Strata programmable valve for shunt-dependent hydrocephalus: the pediatric experience at a single institution.

Authors:  Edward S Ahn; Markus Bookland; Benjamin S Carson; Jon D Weingart; George I Jallo
Journal:  Childs Nerv Syst       Date:  2006-10-07       Impact factor: 1.475

2.  Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

Authors:  F G Shellock; R Habibi; J Knebel
Journal:  AJNR Am J Neuroradiol       Date:  2006-03       Impact factor: 3.825

3.  Magnetic toys: forbidden for pediatric patients with certain programmable shunt valves?

Authors:  Tycho J Zuzak; Bettina Balmer; Daniel Schmidig; Eugen Boltshauser; Michael A Grotzer
Journal:  Childs Nerv Syst       Date:  2008-12-05       Impact factor: 1.475

4.  Worldwide survey of damage from swallowing multiple magnets.

Authors:  Alan E Oestreich
Journal:  Pediatr Radiol       Date:  2008-11-20

5.  Radiation risk due to shunted hydrocephalus and the role of MR imaging-safe programmable valves.

Authors:  S Krishnamurthy; B Schmidt; M D Tichenor
Journal:  AJNR Am J Neuroradiol       Date:  2012-11-01       Impact factor: 3.825

6.  Interactions between programmable shunt valves and the iPad 3 with Smart Cover.

Authors:  Yizheng He; Rory K J Murphy; Jarod L Roland; David D Limbrick
Journal:  Childs Nerv Syst       Date:  2013-02-20       Impact factor: 1.475

Review 7.  Gastrointestinal damage caused by swallowing multiple magnets.

Authors:  Shiqi Liu; Jianhui Li; Yi Lv
Journal:  Front Med       Date:  2012-08-11       Impact factor: 4.592

8.  Management of neonatal hydrocephalus: feasibility of use and safety of two programmable (Sophy and Polaris) valves.

Authors:  Juan F Martínez-Lage; María-José Almagro; Isabel Sanchez Del Rincón; Miguel A Pérez-Espejo; Claudio Piqueras; Raúl Alfaro; Javier Ros de San Pedro
Journal:  Childs Nerv Syst       Date:  2007-10-09       Impact factor: 1.475

9.  Shunt technology for infants and a lifetime.

Authors:  Víctor J Fernández Cornejo; Samer K Elbabaa
Journal:  Childs Nerv Syst       Date:  2021-07-08       Impact factor: 1.475

10.  In vitro hydrodynamic properties of the Miethke ProGAV hydrocephalus shunt.

Authors:  David M Allin; Zofia H Czosnyka; Marek Czosnyka; Hugh K Richards; John D Pickard
Journal:  Cerebrospinal Fluid Res       Date:  2006-06-29
View more