| Literature DB >> 15829118 |
L J Owen1, S Gillingwater, B G Keevil.
Abstract
Prednisolone is a commonly prescribed corticosteroid used in the treatment of many diseases. Despite high doses of prednisolone, some patients appear to have subtherapeutic concentrations of the drug. It would be useful to measure prednisolone in this group to determine if they have poor absorption or compliance. Hence, we have developed a liquid chromatography-tandem mass spectrometry method for the determination of prednisolone in serum. Chromatography was performed using a C18 column, giving a retention time for both prednisolone and deuterated prednisolone (internal standard) of 1.6 min. Two transitions were monitored for both prednisolone and deuterated prednisolone. These were m/z 361.2 > 343.0 and m/z 361.2 > 146.9 for prednisolone, and m/z 367.2 > 349.0 and m/z 367.2 > 149.9 for the internal standard. The intra- and inter-batch imprecision was < 7% in both cases over a concentration range of 62.5-750 microg/L. The imprecision at the lower limit was 8%, the lower limit of quantitation was determined to be 30 microg/L and the method was linear up to 5000 microg/L. The method allows rapid prednisolone analysis because of a simplified sample extraction step, and has a cycle time of 3.5 min.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15829118 DOI: 10.1258/0004563053492810
Source DB: PubMed Journal: Ann Clin Biochem ISSN: 0004-5632 Impact factor: 2.057