Revisiting the accuracy of peak flow meters: a double-blind study using formal methods of agreement.

Z Nazir1, S Razaq, S Mir, M Anwar, G Al Mawlawi, M Sajad, A Shehab, R S Taylor.   

Abstract

BACKGROUND: There is widespread use of peak flow meters in both hospitals and general practice. Previous studies to assess peak flow meter accuracy have shown significant differences in the values obtained from different meters. However, many of these studies did not use human subjects for peak flow measurements and did not compare meters of varying usage. In this study human subjects have been used with meters of varying usage.
METHODS: Participants were tested using two new (meters A and C) and one old peak flow meter (meter B) in random order. The study was double-blinded. Participants were recruited from the university campus.
RESULTS: Four hundred and nine individuals participated. The difference between peak flow means of A and B was -9.93 l/min (95% CI: -12.37 to -7.48, P<0.0001). The difference between peak flow means of B and C was 20.08 l/min (95% CI: 17.85-22.29, P<0.0001). The difference between peak flow means of A and C was 10.15 l/min (95% CI: 7.68-12.61, P<0.0001).
CONCLUSION: There was a significant difference between the values obtained from the new and old peak flow meters and also between the two new peak flow meters. We conclude that there is need for caution in interchangeably using flow meters in clinical practice.

Entities:  

Mesh:

Year:  2004        PMID: 15823456     DOI: 10.1016/j.rmed.2004.10.015

Source DB:  PubMed          Journal:  Respir Med        ISSN: 0954-6111            Impact factor:   3.415


  5 in total

1.  Comparison of five portable peak flow meters.

Authors:  Glaucia Nency Takara; Gualberto Ruas; Bruna Varanda Pessoa; Luciana Kawakami Jamami; Valéria Amorim Pires Di Lorenzo; Mauricio Jamami
Journal:  Clinics (Sao Paulo)       Date:  2010-05       Impact factor: 2.365

2.  The Effect of Training Inhalation Technique with or without Spacer on Maximum Expiratory Flow Rate and Inhaler Usage Skills in Asthmatic Patients: A Randomized Controlled Trial.

Authors:  Hashem Rahmati; Fatemeh Ansarfard; Fariba Ghodsbin; Mohammad Ali Ghayumi; Mehrab Sayadi
Journal:  Int J Community Based Nurs Midwifery       Date:  2014-10

3.  Evaluating the Validity of an Automated Device for Asthma Monitoring for Adolescents: Correlational Design.

Authors:  Hyekyun Rhee; Michael J Belyea; Mark Sterling; Mark F Bocko
Journal:  J Med Internet Res       Date:  2015-10-16       Impact factor: 5.428

4.  Evaluation of a hospital pharmacy-based pharmaceutical care services for asthma patients.

Authors:  Elkhansa Abdelhamid; Abdelmoneim Awad; Abdellatif Gismallah
Journal:  Pharm Pract (Granada)       Date:  2008-03-10

5.  Testing the accuracy of a novel digital peak flow meter aligned with a smartphone app compared to a lab spirometer: A pilot work.

Authors:  Panagiotis Sakkatos; Andrew Williams
Journal:  Digit Health       Date:  2021-05-21
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.