Tailoring high-cut-off membranes and feasible application in sepsis-associated acute renal failure: in vitro studies.

Filippo Mariano1, Valentina Fonsato, Giacomo Lanfranco, Robert Pohlmeier, Claudio Ronco, Giorgio Triolo, Giovanni Camussi, Ciro Tetta, Jutta Passlick-Deetjen.   

Abstract

BACKGROUND: As removal of pro-inflammatory cytokines is limited in conventional diffusive or convective extracorporeal therapies, we studied in two polysulphone membranes with an industrial albumin sieving coefficient of 0.05 (Type A) and 0.13 (Type B) cytokine (IL-6, IL-8, IL-1beta, IL-1ra, TNF-alpha) and plasma protein (albumin, cystatin C, total proteins) permeability profiles. Based on the convective membrane permeability, we evaluated in vitro the dialytic modality that could provide an acceptable balance between high cytokine and low albumin clearances.
METHODS: Cytokine and plasma protein sieving coefficient (SC) and clearance were studied in (i) post-dilutional haemofiltration mode at 20% fixed ultrafiltration rate; (ii) haemodialysis mode (dialysate flow rate of 3 and 5 l/h); and (iii) haemodiafiltration mode (dialysate flow rate of 3 or 5 l/h with 0.5 l/h of ultrafiltrate).
RESULTS: In haemofiltration mode both Type A and Type B haemodialysers at QB 150 ml/min exhibited similar median SC nearly up to 1 for IL-1beta and IL-1ra, at about 0.6 for IL-6, 0.4 for IL-8 and 0.7 for TNF-alpha, with clearance values ranging from 15 to 30 ml/min. SC were independent of blood flow and were stable throughout the whole experiment. Albumin SC was higher in Type B than in Type A and rapidly decreased from 0.2 to 0.02 and from 0.5 to 0.04 within 3 h for haemodialyser Types A and B, respectively. Cytokine SC was lower in haemodialysis than in haemodiafiltration and haemofiltration mode, and by increasing dialysate flow from 3 up to 5 l/h in both haemodialysis and haemodiafiltration mode, SC for all tested cytokines decreased. However, at 5 l/h clearances were not different or were higher, since increased amounts of dialysate outlet compensated for the decreased SC. Albumin clearances in haemodialysis and haemodiafiltration mode after 360 min at 5 l/h were 0.81 and 0.91 ml/min, respectively.
CONCLUSIONS: Our studies show that a mixed convective and diffusive technique ensures high cytokine clearances with an acceptable loss of albumin.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15814541     DOI: 10.1093/ndt/gfh776

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  3 in total

Review 1.  Acute kidney injury in the ICU: from injury to recovery: reports from the 5th Paris International Conference.

Authors:  Rinaldo Bellomo; Claudio Ronco; Ravindra L Mehta; Pierre Asfar; Julie Boisramé-Helms; Michael Darmon; Jean-Luc Diehl; Jacques Duranteau; Eric A J Hoste; Joannes-Boyau Olivier; Matthieu Legrand; Nicolas Lerolle; Manu L N G Malbrain; Johan Mårtensson; Heleen M Oudemans-van Straaten; Jean-Jacques Parienti; Didier Payen; Sophie Perinel; Esther Peters; Peter Pickkers; Eric Rondeau; Miet Schetz; Christophe Vinsonneau; Julia Wendon; Ling Zhang; Pierre-François Laterre
Journal:  Ann Intensive Care       Date:  2017-05-04       Impact factor: 6.925

2.  Effectiveness of High Cut-Off Hemofilters in the Removal of Selected Cytokines in Patients During Septic Shock Accompanied by Acute Kidney Injury-Preliminary Study.

Authors:  Grzegorz Kade; Arkadiusz Lubas; Agnieszka Rzeszotarska; Jolanta Korsak; Stanisław Niemczyk
Journal:  Med Sci Monit       Date:  2016-11-13

Review 3.  Recent evolution of renal replacement therapy in the critically ill patient.

Authors:  Claudio Ronco
Journal:  Crit Care       Date:  2006-02       Impact factor: 9.097

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.