Literature DB >> 15810971

Intracanal temperature rise evaluation during the usage of the System B: replication of intracanal anatomy.

J C Villegas1, T Yoshioka, Ch Kobayashi, H Suda.   

Abstract

AIM: To evaluate and determine intracanal temperature rises at 2 and 4 mm from the working length (WL) necessary to obtain proper replication of intracanal anatomy with gutta-percha (GP) using the System B heat source during vertical condensation.
METHODOLOGY: A split-tooth model was prepared and artificial shallow depressions were cut in the buccal canal wall 2 and 4 mm from the WL. At the same level on the palatal wall holes were drilled to adapt two thermocouples. The canal was filled using GP in a vertical condensation technique by placing the System B plugger at 2 and 4 mm from the WL in groups A and B, respectively. Two control groups in which no GP was used were carried out placing the plugger 2 and 4 mm from the WL (groups A.c and B.c, respectively) and activating the heat source. Recording of temperature rise was carried out during the filling procedure for groups A and B and during activation for control groups A.c and B.c; the highest temperatures were recorded. After each filling was completed, 3 min were allowed for the GP to cool and the model divided to reveal the filling. Images of the GP were taken with a CCD camera to evaluate the presence of replication of artificial round depressions. In control groups, the temperature was recorded for 20 s after a 3 s activation of the heat source. The rise in temperature was compared between the groups individually at each level (2 or 4 mm) and statistically analysed using one-way anova and Fisher PLSD tests at 5% of significance level (P < 0.05).
RESULTS: Mean temperature rises of 14 +/- 3 and 12 +/- 2 degrees C at 2 and 4 mm from the WL, respectively, were observed in group A fillings, and 4 +/- 1 and 6 +/- 1 degrees C at 2 and 4 mm, respectively, in group B fillings. Recordings at 2 mm showed significantly (P < 0.05) higher temperature rises with group A.c when compared with groups B and B.c. Replication of intracanal anatomy with GP was always found in group A fillings at both levels but only 4 mm from the WL in group B fillings.
CONCLUSIONS: Positioning the plugger close to WL and a temperature rise of 6 degrees C were necessary to obtain replication of intracanal anatomy. A mean temperature rise of 4 degrees C at 2 mm from WL (group B) resulted in no replication of intracanal anatomy. Further studies simulating clinical conditions are necessary.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15810971     DOI: 10.1111/j.1365-2591.2005.00934.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  3 in total

1.  Thermal analysis and stability of commercially available endodontic obturation materials.

Authors:  H W Roberts; T C Kirkpatrick; B E Bergeron
Journal:  Clin Oral Investig       Date:  2017-02-10       Impact factor: 3.573

2.  System B, Endo-Twinn and E-Fill. True temperatures inside the canal.

Authors:  Enzo Cumbo; Riccardo Russo; Giuseppe Gallina
Journal:  J Conserv Dent       Date:  2012-10

3.  Plugger temperature of cordless heat carriers according to the time elapsed.

Authors:  Hoon-Sang Chang; Se-Hee Park; Kyung-Mo Cho; Jin-Woo Kim
Journal:  Restor Dent Endod       Date:  2018-02-07
  3 in total