Literature DB >> 15810791

Power-line interference cancellation in ECG signals.

I Dotsinsky1, T Stoyanov.   

Abstract

The efficiency of notchfilters and a subtraction procedure for power-line interference cancellation in electrocardiogram (ECG) signals is assessed. In contrast with the subtraction procedure, widely used digital notch filters unacceptably affect QRS complexes. The procedure eliminates interferences of variable amplitude and frequency. The frequency modulations are overcome by adaptive synchronized sampling. Initially, this is accomplished by current hardware power-line frequency measurement. Because this approach is impossible in battery-supplied and some computer-aidd devices, a software measurement of the power-line interference period is developed.

Entities:  

Mesh:

Year:  2005        PMID: 15810791     DOI: 10.2345/0899-8205(2005)39[155:PICIES]2.0.CO;2

Source DB:  PubMed          Journal:  Biomed Instrum Technol        ISSN: 0899-8205


  5 in total

1.  A wearable mobihealth care system supporting real-time diagnosis and alarm.

Authors:  J W Zheng; Z B Zhang; T H Wu; Y Zhang
Journal:  Med Biol Eng Comput       Date:  2007-07-06       Impact factor: 2.602

2.  Approximating ideal filters by systems of fractional order.

Authors:  Ming Li
Journal:  Comput Math Methods Med       Date:  2012-01-16       Impact factor: 2.238

3.  Removal of power-line interference from the ECG: a review of the subtraction procedure.

Authors:  Chavdar Levkov; Georgy Mihov; Ratcho Ivanov; Ivan Daskalov; Ivaylo Christov; Ivan Dotsinsky
Journal:  Biomed Eng Online       Date:  2005-08-23       Impact factor: 2.819

4.  Suppression of AC railway power-line interference in ECG signals recorded by public access defibrillators.

Authors:  Ivan Dotsinsky
Journal:  Biomed Eng Online       Date:  2005-11-26       Impact factor: 2.819

5.  A hybrid approach to the simultaneous eliminating of power-line interference and associated ringing artifacts in electrocardiograms.

Authors:  Xiaolin Zhou; Yuanting Zhang
Journal:  Biomed Eng Online       Date:  2013-05-14       Impact factor: 2.819

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.