Literature DB >> 1580376

Determination of plutonium-239 + plutonium-240 and plutonium-241 in environmental samples using low-level liquid scintillation spectrometry.

Y F Yu1, H E Bjørnstad, B Salbu.   

Abstract

A radiochemical method for the simultaneous determination of 239Pu + 240Pu and 241Pu in environmental samples has been developed. In the course of the analysis a 236Pu tracer was used for estimating the chemical yield of plutonium isotopes. After suitable pre-treatment of the sample, the plutonium nuclides in solution were coprecipitated with iron(III) hydroxide and calcium oxalate and isolated further from impurities and interfering radionuclides by means of anion-exchange chromatography. Plutonium isotopes in the eluate (NH4I-HCI) were converted into nitrate form and then extracted with 20 ml of 5% bis(2-ethylhexyl) hydrogen phosphate extractive cocktail. The final organic solution was measured spectrometrically using an ultra-low-level liquid scintillation spectrometer, Quantulus (LKB, 1220 Wallac). The chemical yields of plutonium range from 25 to 50% for 100 I of sea-water and 30 to 60% for 40 g of dried soil sample. The counting efficiencies are nearly 100% for 239Pu + 240Pu and 48.8% for 241Pu, respectively. The detection limits were estimated to be 0.20 mBq for 239Pu + 240Pu and 2.2 mBq for 241Pu, respectively. The proposed procedure has been tested for the simultaneous determination of 239Pu + 240Pu and 241Pu in sea-water (Irish Sea, North Sea) and soils (Cumbrian coast, UK; Byelorussia, USSR).

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1580376     DOI: 10.1039/an9921700439

Source DB:  PubMed          Journal:  Analyst        ISSN: 0003-2654            Impact factor:   4.616


  1 in total

1.  Rapid method for plutonium-241 determination in soil samples.

Authors:  M Piekarz; A Komosa
Journal:  J Radioanal Nucl Chem       Date:  2014-01-24       Impact factor: 1.371

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.