Abstract
The dynamic equilibrium exploited by balanced steady-state free precession imaging develops slowly because its formation is dependent on both spin-spin and spin-lattice relaxation times. Attempting to image before steady state is established results in artifacts due to transient signal oscillations. Using a starter sequence to precondition the spin system can significantly reduce the delay before imaging. An improved design for a steady-state starter sequence is presented. The new sequence has the advantage of uniformly exciting the steady-state response for all resonance offsets and can be phase cycled to suppress banding artifacts. Copyright 2005 Wiley-Liss, Inc.Mesh:
Year: 2005 PMID: 15799066 DOI: 10.1002/mrm.20437
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668