| Literature DB >> 15797799 |
Phensri Thongnopnua1, Chotima Poeaknapo.
Abstract
A high-performance liquid chromatographic method for indirect determination of enalapril in human plasma, was developed and validated. An exogenous angiotensin converting enzyme after drug inhibition was determined by reacting with hippuryl-histidyl-leucine to produce hippuric acid (HA) which was inversely proportional to the amount of enalaprilat in plasma. The HPLC was carried out on a Lichrosphere 60RP-select B, C18, 5 microm (125 mm x 4.0 mm i.d.) column at flow rate of 1.0 ml/min. The analysis time per injection was within 6.5 min. The lowest concentration of enalaprilat to be quantitated was 3.0 ng/ml with the acceptable accuracy and precision. This successfully developed method was practically and accurately used for pharmacokinetics and bioequivalent study of enalapril.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15797799 DOI: 10.1016/j.jpba.2004.10.053