| Literature DB >> 15797361 |
W C Yam1, K H Chan, K H Chow, L L M Poon, H Y Lam, K Y Yuen, W H Seto, J S M Peiris.
Abstract
BACKGROUND: The protocols of WHO network laboratories facilitated development of rapid diagnosis for SARS coronavirus (CoV) using reverse transcription (RT)-PCR assays. However, several reports have shown that conventional and real-time PCR assays were very specific for SARS CoV but lack sensitivity depending on the assay, specimen, and time course of disease.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15797361 PMCID: PMC7108323 DOI: 10.1016/j.jcv.2004.09.029
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
Performance of real-time PCR assays for rapid diagnosis of SARS CoV infection
| Nature of specimens (no.) | No. tested | Sero-conversion | No. positive (range of RNA copies detected/reaction) | ||
|---|---|---|---|---|---|
| WHO-HKU | Artus | Roche | |||
| Epidemic period (April–May 2003) | |||||
| Respiratory specimens (64) | 34 | + | 34 | 34 (0.48–2 × 106) | 34 (7–1 × 108) |
| 13 | + | 0 | 7 (0.13–3 × 104) + [1] | 6 (3–7 × 103) + [1] | |
| 17 | − | 0 | 0 + [3] | 0 + [2] | |
| Stool (44) | 23 | + | 23 | 23 (0.25–1.7 × 106) | 23 (17–4 × 107) |
| 9 | + | 0 | 6 (1.8–3.7) + [2] | 5 (0.44–43.9) + [2] | |
| 1 | − | 0 | 1 (1.3) | 0 | |
| 11 | − | 0 | 0 + [3] | 0 + [2] | |
| Urine (29) | 13 | + | 13 | 13 (2.3–1.6 × 104) | 13 (3.9–2 × 104) |
| 9 | + | 0 | 5 (2.7–3.7 × 103) + [1] | 5 (72–102) + [1] | |
| 7 | − | 0 | 0 + [1] | 0 + [1] | |
| Controls | |||||
| Respiratory specimens | 9 | ND | 0 | 0 | 0 |
| Post-epidemic period (August–September 2003) | |||||
| Respiratory specimens | 56 | − | 0 | 0 + [5] | 0 + [5] |
A four-fold rise or more in antibody titer against coronavirus.
Values in [ ] indicate number of samples showing presence of PCR inhibitors.
Samples positive for other viral pathogens included 4 flu A, 2 adenovirus, and 3 RSV by immunofluorescence (Chan et al., 2002).
Samples positive for other viral pathogens included 8 flu A, 1 adenovirus, and 23 RSV by immunofluorescence (Chan et al., 2002).
Fig. 1Comparison of real-time PCR assays for detection of SARS CoV.