Literature DB >> 15794563

High performance liquid chromatography equipped with a cathodic detector and column-switching device as a high-throughput method for a phosphatase assay with p-nitrophenyl phosphate.

Yuji Yamauchi1, Megumi Ido, Hatsuo Maeda.   

Abstract

LC coupled to an electrochemical detector (LC-EC) operating in cathodic mode with a column-switching system realizes a high-throughput detection of p-nitrophenol (NP). The measurement-time for each NP sample was shortened to 20 s, and the successive analyses of 39 samples was completed within 13 min. In the present system, the limits of detection and quantification were 0.15 and 0.20 microM, respectively, and further, up to 25 microM, a linear calibration curve was afforded. Relative standard deviations for standard solutions of 0.20, 1.0, and 25 microM NP were 4.3, 2.0, and 1.1% (n = 5), respectively. Between-run precisions of the analysis of 5.0 and 25 microM NP over 6 days were 4.8 and 1.3%, respectively. A comparison with the commonly used Bessey-Lowry-Brock method indicates that the present LC-EC is useful for the high-throughput assay of acid and alkaline phosphatases in urine and blood samples with a p-nitrophenyl phosphate substrate.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15794563     DOI: 10.1016/j.chroma.2005.01.071

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  A dual-signal fluorescent probe for detection of acid phosphatase.

Authors:  Shuangqin Li; Guoqing Fu; Yaya Wang; Yueci Xiang; Shuai Mu; Yixuan Xu; Xiaoyan Liu; Haixia Zhang
Journal:  Anal Bioanal Chem       Date:  2021-05-01       Impact factor: 4.142

2.  A fluorescent and colorimetric dual-channel sensor based on acid phosphatase-triggered blocking of internal filtration effect.

Authors:  Fanpeng Ran; Chen Ma; Yueci Xiang; Yixuan Xu; Xiaoyan Liu; Haixia Zhang
Journal:  Mikrochim Acta       Date:  2021-08-02       Impact factor: 5.833