Performance evaluation of the application of body fluids on the Sysmex XE-2100 series automated hematology analyzer.

Lesley Kresie1, Diana Benavides, Pam Bollinger, Jerelyn Walters, Deanna Pierson, Terri Richmond, Kelly Issa-Dyer, Michelle Fahs.   

Abstract

Body fluid analysis on the Sysmex XE-2100 series automated hematology analyzer was evaluated at 4 hospitals (Baylor University Medical Center, Dallas, TX; St. John's Mercy, St. Louis, MO; Carle Clinic, Urbana, IL; and ACL Laboratories, West Allis, WI, USA). The total nucleated cell and red blood cell (RBC) counts of 493 samples were obtained with the Sysmex XE-2100 automated hematology analyzer and compared with results obtained by manual chamber counting. Seventy-eight samples were not suitable for evaluation because of the presence of clots, crystals, error messages related to white blood cell (WBC) and RBC parameters, and so on. Pearson correlation coefficients for the WBC parameter were 0.99 for cerebrospinal fluid, 0.95 for serous fluid, 0.99 for synovial fluid, and 0.99 for samples of combined body fluids. Ninety-six samples were used to compare RBC counting methods because these samples had RBC counts greater than 0.01 10 6/mL. The Pearson correlation coefficients for the RBC parameter were 0.96 for cerebrospinal fluid, 0.97 for serous fluid, 0.97 for synovial fluid, and 0.97 for samples of combined body fluids. Carryover, precision, and linearity studies also performed for WBC and RBC counts yielded very good results.

Mesh:

Year:  2005        PMID: 15790549     DOI: 10.1532/LH96.04067

Source DB:  PubMed          Journal:  Lab Hematol        ISSN: 1080-2924


  5 in total

1.  Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i.

Authors:  Sabrina Buoro; Sara Apassiti Esposito; MariaGrazia Alessio; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Spurious hyperglycaemia impairs automated leucocyte counting. A pilot study with two different haematological analysers.

Authors:  Ruggero Buonocore; Alessandra Picanza; Daniela Gennari; Silvia Pipitone; Giuseppe Lippi
Journal:  Blood Transfus       Date:  2015-10       Impact factor: 3.443

3.  Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion.

Authors:  Weiyi Xu; Qian Yu; Lixia Xie; Baode Chen; Ling Zhang
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

4.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

5.  Analytical validation of the Sysmex XT-2000iV for cell counts in canine and feline effusions and concordance with cytologic diagnosis.

Authors:  Nazaré Pinto da Cunha; Alessia Giordano; Mario Caniatti; Saverio Paltrinieri
Journal:  Vet Clin Pathol       Date:  2009-02-02       Impact factor: 1.180

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.