Use of a multi-stage exercise test to assess the responsiveness of rate-adaptive pacemakers in dogs.

L Ferasin1, M Faena, S M Henderson, K Langford, P G G Darke.   

Abstract

OBJECTIVES: To assess the ventricular rate response of rate-adaptive (VVIR) pacemakers in dogs using a multi-stage exercise test.
METHODS: The rate-responsiveness of VVIR pacemakers was assessed in seven dogs with complete atrioventricular (AV) block and implanted with various models of pulse generators (six motion sensors and one automatic dual-sensor rate-response pacemaker). Response activity was assessed with a multi-stage exercise test on a treadmill. Atrial and ventricular rate were analysed retrospectively at the end of the test and the AV ratio was calculated after each minute of exercise.
RESULTS: During exercise, the mean (sd) AV ratio recorded in all paced dogs was 1.7 (0.5) (expected physiological ratio 1.0), although a variety of individual performances was observed. A poor response (AV ratio 2.8 [0.2]) was obtained with the automatic dual-sensor pacemaker, suggesting that this type of rate-responsive device may not be indicated for implantation in dogs with complete AV block. The overall AV ratio for the six dogs implanted with motion sensors was 1.4 (0.2), showing a better performance of these pacemakers during exercise. CLINICAL SIGNIFICANCE: This multi-stage exercise test represents an easy and repeatable method for assessing the accuracy of rate-responsive sensors and offers valuable information for the correct setting of VVIR pacemakers in dogs.

Entities:  

Mesh:

Year:  2005        PMID: 15789806     DOI: 10.1111/j.1748-5827.2005.tb00301.x

Source DB:  PubMed          Journal:  J Small Anim Pract        ISSN: 0022-4510            Impact factor:   1.522


  3 in total

1.  Reliability of an incremental exercise test to evaluate acute blood lactate, heart rate and body temperature responses in Labrador retrievers.

Authors:  Luca Ferasin; Samuele Marcora
Journal:  J Comp Physiol B       Date:  2009-05-20       Impact factor: 2.200

2.  Successful application of closed loop stimulation pacemakers with remote monitoring in 3 miniature donkeys with syncope.

Authors:  Lisa De Lange; Glenn Van Steenkiste; Ingrid Vernemmen; Lisse Vera; Kristel M C Cromheeke; Ulla Walser; Hans Meert; Annelies Decloedt; Gunther van Loon
Journal:  J Vet Intern Med       Date:  2021-11-06       Impact factor: 3.333

Review 3.  Indications for permanent pacing in dogs and cats.

Authors:  R A Santilli; F Giacomazzi; D M Porteiro Vázquez; M Perego
Journal:  J Vet Cardiol       Date:  2019-01-29       Impact factor: 1.701

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.