Physical characterization of a prototype selenium-based full field digital mammography detector.

Robert S Saunders1, Ehsan Samei, Jonathan L Jesneck, Joseph Y Lo.   

Abstract

The purpose of this study was to measure experimentally the physical performance of a prototype mammographic imager based on a direct detection, flat-panel array design employing an amorphous selenium converter with 70 microm pixels. The system was characterized for two different anode types, a molybdenum target with molybdenum filtration (Mo/Mo) and a tungsten target with rhodium filtration (W/Rh), at two different energies, 28 and 35 kVp, with approximately 2 mm added aluminum filtration. To measure the resolution, the presampled modulation transfer function (MTF) was measured using an edge method. The normalized noise power spectrum (NNPS) was measured by two-dimensional Fourier analysis of uniformly exposed mammograms. The detective quantum efficiencies (DQEs) were computed from the MTFs, the NNPSs, and theoretical ideal signal to noise ratios. The MTF was found to be close to its ideal limit and reached 0.2 at 11.8 mm(-1) and 0.1 at 14.1 mm(-1) for images acquired at an RQA-M2 technique (Mo/Mo anode, 28 kVp, 2 mm Al). Using a tungsten technique (MW2; W/Rh anode, 28 kVp, 2 mm Al), the MTF went to 0.2 at 11.2 mm(-1) and to 0.1 at 13.3 mm(-1). The DQE reached a maximum value of 54% at 1.35 mm(-1) for the RQA-M2 technique at 1.6 microC/kg and achieved a peak value of 64% at 1.75 mm(-1) for the tungsten technique (MW2) at 1.9 microC/kg. Nevertheless, the DQE showed strong exposure and frequency dependencies. The results indicated that the detector offered high MTFs and DQEs, but structured noise effects may require improved calibration before clinical implementation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15789606     DOI: 10.1118/1.1855033

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  A technique optimization protocol and the potential for dose reduction in digital mammography.

Authors:  Nicole T Ranger; Joseph Y Lo; Ehsan Samei
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system.

Authors:  Paula Toroi; Federica Zanca; Kenneth C Young; Chantal van Ongeval; Guy Marchal; Hilde Bosmans
Journal:  Eur Radiol       Date:  2007-02-01       Impact factor: 5.315

Review 3.  Flat-detector computed tomography (FD-CT).

Authors:  Willi A Kalender; Yiannis Kyriakou
Journal:  Eur Radiol       Date:  2007-06-23       Impact factor: 5.315

4.  Detection of simulated microcalcifications in a phantom with digital mammography: effect of pixel size.

Authors:  Sankararaman Suryanarayanan; Andrew Karellas; Srinivasan Vedantham; Ioannis Sechopoulos; Carl J D'Orsi
Journal:  Radiology       Date:  2007-05-23       Impact factor: 11.105

5.  A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems.

Authors:  G K Yadava; A T Kuhls-Gilcrist; S Rudin; V K Patel; K R Hoffmann; D R Bednarek
Journal:  Phys Med Biol       Date:  2008-08-22       Impact factor: 3.609

6.  Imaging performance of an amorphous selenium digital mammography detector in a breast tomosynthesis system.

Authors:  Bo Zhao; Wei Zhao
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 7.  [Basic principles of flat detector computed tomography (FD-CT)].

Authors:  Y Kyriakou; T Struffert; A Dörfler; W A Kalender
Journal:  Radiologe       Date:  2009-09       Impact factor: 0.635

8.  Investigation of noise sources for digital radiography systems.

Authors:  Lutfi Ergun; Turan Olgar
Journal:  Radiol Phys Technol       Date:  2016-10-01

9.  Task-based strategy for optimized contrast enhanced breast imaging: analysis of six imaging techniques for mammography and tomosynthesis.

Authors:  Lynda C Ikejimba; Nooshin Kiarashi; Sujata V Ghate; Ehsan Samei; Joseph Y Lo
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

Review 10.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.