Performance comparison of an active matrix flat panel imager, computed radiography system, and a screen-film system at four standard radiation qualities.

P Monnin1, D Gutierrez, S Bulling, D Lepori, J F Valley, F R Verdun.   

Abstract

Four standard radiation qualities (from RQA 3 to RQA 9) were used to compare the imaging performance of a computed radiography (CR) system (general purpose and high resolution phosphor plates of a Kodak CR 9000 system), a selenium-based direct flat panel detector (Kodak Direct View DR 9000), and a conventional screen-film system (Kodak T-MAT L/RA film with a 3M Trimax Regular screen of speed 400) in conventional radiography. Reference exposure levels were chosen according to the manufacturer's recommendations to be representative of clinical practice (exposure index of 1700 for digital systems and a film optical density of 1.4). With the exception of the RQA 3 beam quality, the exposure levels needed to produce a mean digital signal of 1700 were higher than those needed to obtain a mean film optical density of 1.4. In spite of intense developments in the field of digital detectors, screen-film systems are still very efficient detectors for most of the beam qualities used in radiology. An important outcome of this study is the behavior of the detective quantum efficiency of the digital radiography (DR) system as a function of beam energy. The practice of users to increase beam energy when switching from a screen-film system to a CR system, in order to improve the compromise between patient dose and image quality, might not be appropriate when switching from screen-film to selenium-based DR systems.

Entities:  

Mesh:

Year:  2005        PMID: 15789578     DOI: 10.1118/1.1843451

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

Review 1.  Application of QC_DR software for acceptance testing and routine quality control of direct digital radiography systems: initial experiences using the Italian Association of Physicist in Medicine quality control protocol.

Authors:  Andrea Nitrosi; Marco Bertolini; Giovanni Borasi; Andrea Botti; Adriana Barani; Stefano Rivetti; Luisa Pierotti
Journal:  J Digit Imaging       Date:  2008-09-03       Impact factor: 4.056

2.  Effect of X-ray beam quality on determination of exposure index.

Authors:  Shogo Yasumatsu; Nobukazu Tanaka; Kensuke Iwase; Yoichiro Shimizu; Junji Morishita
Journal:  Radiol Phys Technol       Date:  2015-11-30

3.  SAPHIRE (scintillator avalanche photoconductor with high resolution emitter readout) for low dose x-ray imaging: spatial resolution.

Authors:  Dan Li; Wei Zhao
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071