Literature DB >> 15787920

Umbilical vein catheterization under electrocardiogram guidance.

Ban C H Tsui1, Gareth J Richards, John Van Aerde.   

Abstract

BACKGROUND: In the neonate, umbilical venous catheters (UVC) are inserted and advanced blindly to a predetermined length from the umbilicus. The reported rates for UVC misplacement into the liver (and occasionally the spleen) range from 20 to 37%. Radiographs are routinely used to confirm the positioning of UVCs. This involves movement of often critically ill infants, as well as radiation exposure. This pilot study examines the potential value of confirming UVC placement in neonates using ECG.
METHODS: In critically ill neonates, a conductive Johans ECG adapter was connected to a UVC. A satisfactory tracing (lead II) was obtained (right arm lead connected to the adapter) when the UVC was filled with saline solution allowing the catheter tip to become a unipolar ECG electrode. The UVC was then advanced from the umbilicus until the tip reached the inferior vena cava (IVC) within the thoracic region, as demonstrated by appearance of normal sized QRS complexes with small P-waves. A small QRS indicated the catheter was below the diaphragm. The appearance of a tall positive P-wave indicated the tip was at the right atrium level. The UVC was then withdrawn until the P-wave size returned to normal. The final UVC position was later confirmed by X-ray.
RESULTS: Eight neonates were studied. The figure shows typical ECG tracings when the UVC was placed in the liver, IVC, and right atrium, respectively. Three malpositioned catheters were detected (2 into liver and 1 into spleen).
CONCLUSIONS: Based on these cases, the insertion of UVCs in neonates can be guided with ECG by observing sequential and characteristic alterations in P-waves and QRS complexes, thereby reducing the use of X-rays. In addition, this technique could prove to be beneficial in remote healthcare facilities where X-ray machines may not be readily available and quick intravenous access is required to transport sick neonates to major centers.

Entities:  

Mesh:

Year:  2005        PMID: 15787920     DOI: 10.1111/j.1460-9592.2005.01433.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  2 in total

1.  [Arterial and central venous catheters in neonates and infants].

Authors:  M Stocker; T M Berger
Journal:  Anaesthesist       Date:  2006-08       Impact factor: 1.041

2.  An Accuracy Study of the Intracavitary Electrocardiogram (IC-ECG) Guided Peripherally Inserted Central Catheter Tip Placement among Neonates.

Authors:  Lian-Juan Zhou; Hong-Zhen Xua; Mei-Fang Xu; Yan Hu; Xiao-Fang Lou
Journal:  Open Med (Wars)       Date:  2017-05-22