Evaluation of the monocyte counting by two automated haematology analysers compared with flow cytometry.

E Grimaldi1, P Carandente, F Scopacasa, M F Romano, M Pellegrino, R Bisogni, M De Caterina.   

Abstract

The aim is to determine the monocyte count performance of the Bayer Diagnostics ADVIA120 and Coulter LH 750 automated haematology analysers and the results obtained by these two instruments were compared with those provided by Becton Dickinson FACScan flow cytometer using the combination of CD45/CD14 MoAb. Linearity and imprecision were also evaluated. The linearity of both instruments was good. Coulter LH 750 showed better precision (4.3%) than ADVIA 120 (9.0%) both within and between batch. A significant correlation (r = 0.973) was found between the LH 750 and the flow cytometry method, while a modest one was observed between the latter and the ADVIA 120 (r = 0.880). When comparing the percentage of monocytes by means of one-way anova and Tukey test, it was found that the LH 750 provided the closest results in comparison with flow cytometry, with no statistical difference between the means (mean difference MO% = 0.6); however the difference was statistically different between the ADVIA 120 and flow cytometry (mean difference MO% = -4.06). These data were confirmed by Altman-Bland and Deming regression analyses.

Entities:  

Mesh:

Year:  2005        PMID: 15784123     DOI: 10.1111/j.1365-2257.2005.00676.x

Source DB:  PubMed          Journal:  Clin Lab Haematol        ISSN: 0141-9854


  3 in total

1.  Properties of human blood monocytes. I. CD91 expression and log orthogonal light scatter provide a robust method to identify monocytes that is more accurate than CD14 expression.

Authors:  Dorothy Hudig; Kenneth W Hunter; W John Diamond; Doug Redelman
Journal:  Cytometry B Clin Cytom       Date:  2013-10-08       Impact factor: 3.058

2.  High Monocyte Count Associated with Human Cytomegalovirus Replication In Vivo and Glucocorticoid Therapy May Be a Hallmark of Disease.

Authors:  Przemyslaw Zdziarski; Andrzej Gamian
Journal:  Int J Mol Sci       Date:  2022-08-24       Impact factor: 6.208

3.  Thrombocytopenia Complicating Transcatheter Aortic Valve Implantation: Differences Between Two New-Generation Devices.

Authors:  Nicola Corcione; Simona Romano; Alberto Morello; Paolo Ferraro; Michele Cimmino; Michele Albanese; Martina Tufano; Daniela Capasso; Salvatore Buonpane; Salvatore Giordano; Martino Pepe; Giuseppe Biondi-Zoccai; Maria Fiammetta Romano; Arturo Giordano
Journal:  J Cardiovasc Transl Res       Date:  2021-03-15       Impact factor: 4.132

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.