Camila Savian1, Pamela Chan, Jennifer Paratz.
Abstract
Including positive end-expiratory pressure (PEEP) in the manual resuscitation bag (MRB) may render manual hyperinflation (MHI) ineffective as a secretion maneuver technique in mechanically ventilated patients. In this study we aimed to determine the effect of increased PEEP or decreased compliance on peak expiratory flow rate (PEF) during MHI. A blinded, randomized study was performed on a lung simulator by 10 physiotherapists experienced in MHI and intensive care practice. PEEP levels of 0-15 cm H(2)O, compliance levels of 0.05 and 0.02 L/cm H(2)O, and MRB type were randomized. The Mapleson-C MRB generated significantly higher PEF (P < 0.01, d = 2.72) when compared with the Laerdal MRB for all levels of PEEP. In normal compliance (0.05 L/cm H(2)O) there was a significant decrease in PEF (P < 0.01, d = 1.45) for a PEEP more than 10 cm H(2)O in the Mapleson-C circuit. The Laerdal MRB at PEEP levels of more than 10 cm H(2)O did not generate a PEF that is theoretically capable of producing two-phase gas-liquid flow and, consequently, mobilizing pulmonary secretions. If MHI is indicated as a result of mucous plugging, the Mapleson-C MRB may be the most effective method of secretion mobilization.Entities:
Mesh:
Year: 2005 PMID: 15781530 DOI: 10.1213/01.ANE.0000147505.98565.AC
Source DB: PubMed Journal: Anesth Analg ISSN: 0003-2999 Impact factor: 5.108