Literature DB >> 15781521

Non-zero basal oxygen flow a hazard to anesthesia breathing circuit leak test.

Joho Tokumine1, Kazuhiro Sugahara, Kouji Gushiken, Minoru Ohta, Tomoaki Matsuyama, Satoko Saikawa.   

Abstract

The non-zero basal flow (BF) of oxygen in anesthesia machines has been set to supply the basal metabolic requirement of oxygen. However, there is no scientific evidence of its necessity. In this study we sought to clarify whether non-zero BF affects leak detection during preanesthetic inspections. Twenty-five participants performed leak tests on anesthesia machines to detect breathing circuit leaks. Artificial leak-producing devices were used to create leaks from 0 to 1.0 L/min. The investigator randomly chose the leak device and connected it into the breathing circuit. Participants, blinded as to the presence or the type of leak producing device, then tested the breathing circuit for leaks. The conventional breathing system leak test was performed with and without BF. The results of leak detection in each leak procedure were analyzed statistically. The leak detection rate of leak test with BF was less than without BF (P < 0.01). We demonstrated that non-zero BF of oxygen decreases the leak detection rate and is an obstacle for leak detection, especially for small leaks. Therefore, we recommend that breathing circuit leak tests should be performed in the absence of BF of oxygen.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15781521     DOI: 10.1213/01.ANE.0000147666.28051.BB

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

1.  Aberrant halt of syringe pump motion: an improved system to prevent false setting of the syringe.

Authors:  Joho Tokumine; Kazuhiro Sugahara; Kenichi Nitta; Tatsuya Fuchigami; Masanori Abe; Kouji Gushiken; Masami Oda; Haruka Okayama
Journal:  J Anesth       Date:  2006       Impact factor: 2.078

2.  Abnormal function of a normal ventilator.

Authors:  Rakesh Garg; Girija P Rath
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2011-04

Review 3.  Anaesthesia machine: checklist, hazards, scavenging.

Authors:  Umesh Goneppanavar; Manjunath Prabhu
Journal:  Indian J Anaesth       Date:  2013-09