Literature DB >> 15773630

Quantitative analysis of synchrotron radiation intravenous angiographic images.

Anna Sarnelli1, Christian Nemoz, Hélène Elleaume, François Estève, Bernard Bertrand, Alberto Bravin.   

Abstract

A medical research protocol on clinical intravenous coronary angiography has been completed at the European Synchrotron Radiation Facility (ESRF) biomedical beamline. The aim was to investigate the accuracy of intravenous coronary angiography based on the K-edge digital subtraction technique for the detection of in-stent restenosis. For each patient, diagnosis has been performed on the synchrotron radiation images and monitored with the conventional selective coronary angiography method taken as the golden standard. In this paper, the methods of image processing and the results of the quantitative analysis are described. Image processing includes beam harmonic contamination correction, spatial deconvolution and the extraction of a 'contrast' and a 'tissue' image from each couple of radiograms simultaneously acquired at energies bracketing the K-edge of iodine. Quantitative analysis includes the estimation of the vessel diameter, the calculation of the absolute iodine concentration profiles along the coronary arteries and the stenosis degree measurement.

Entities:  

Mesh:

Year:  2005        PMID: 15773630     DOI: 10.1088/0031-9155/50/4/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Intravascular Neural Interface with Nanowire Electrode.

Authors:  Hirobumi Watanabe; Hirokazu Takahashi; Masayuki Nakao; Kerry Walton; Rodolfo R Llinás
Journal:  Electron Commun Jpn       Date:  2009-07       Impact factor: 0.324

  1 in total