Literature DB >> 15773627

Consistency test of the electron transport algorithm in the GEANT4 Monte Carlo code.

Emily Poon1, Jan Seuntjens, Frank Verhaegen.   

Abstract

In this work, the condensed history algorithm in GEANT4 (version 4.6.2.p01) is examined. We performed simulations of an ionization chamber composed of water for 1.25 MeV incident photon beams under Fano conditions, and evaluated the consistency of the cavity response for several combinations of electron transport parameters. GEANT4 permits electrons to reach geometric boundaries in large steps, and underestimates lateral displacement near interfaces. Step size artefacts due to distortions in electron fluence and angular distributions reduce the cavity dose by up to 39%. Accurate cavity response can be achieved using severe user-imposed step size restrictions. We suggest that improvements in the electron transport algorithm in GEANT4 should address the handling of boundary crossing.

Entities:  

Mesh:

Year:  2005        PMID: 15773627     DOI: 10.1088/0031-9155/50/4/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Monte Carlo calculations for absolute dosimetry to determine machine outputs for proton therapy fields.

Authors:  Harald Paganetti
Journal:  Phys Med Biol       Date:  2006-05-17       Impact factor: 3.609

2.  Report on G4-Med, a Geant4 benchmarking system for medical physics applications developed by the Geant4 Medical Simulation Benchmarking Group.

Authors:  P Arce; D Bolst; M-C Bordage; J M C Brown; P Cirrone; M A Cortés-Giraldo; D Cutajar; G Cuttone; L Desorgher; P Dondero; A Dotti; B Faddegon; C Fedon; S Guatelli; S Incerti; V Ivanchenko; D Konstantinov; I Kyriakou; G Latyshev; A Le; C Mancini-Terracciano; M Maire; A Mantero; M Novak; C Omachi; L Pandola; A Perales; Y Perrot; G Petringa; J M Quesada; J Ramos-Méndez; F Romano; A B Rosenfeld; L G Sarmiento; D Sakata; T Sasaki; I Sechopoulos; E C Simpson; T Toshito; D H Wright
Journal:  Med Phys       Date:  2020-12-12       Impact factor: 4.071

Review 3.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

4.  Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission.

Authors:  Samir Didi; Abdelilah Moussa; Tayalati Yahya; Zerfaoui Mustafa
Journal:  J Med Phys       Date:  2015 Jul-Sep
  4 in total