Yurika Ikeda-Dantsuji1, Ichiro Konomi1, Ariaki Nagayama1.
Abstract
The Gen-Probe APTIMA Combo 2 assay has previously been reported to have a high sensitivity. The end point of this assay was evaluated using highly purified chlamydial elementary bodies (EBs). The performance of the APTIMA Combo 2 assay was compared with a commercially available PCR kit, AMPLICOR Chlamydia trachomatis. The number of inclusions of C. trachomatis at the end point of the APTIMA Combo 2 assay was 0.005 inclusion-forming units (i.f.u.) ml(-1), which was equivalent to 0.008 EBs per assay. The end point of the AMPLICOR kit was 5 i.f.u. ml(-1) (equivalent to 0.5 EB per assay). The efficacy of the AMPLICOR C. trachomatis assay was inhibited by phosphate or Fe2+ ion, while these had no effect on the APTIMA Combo 2 assay. In conclusion, the APTIMA Combo 2 assay appears to have a greater sensitivity than the AMPLICOR C. trachomatis assay for detection of C. trachomatis, while demonstrating few problems with inhibitory substances such as phosphate and Fe2+ ion.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15770020 DOI: 10.1099/jmm.0.45911-0
Source DB: PubMed Journal: J Med Microbiol ISSN: 0022-2615 Impact factor: 2.472