Jian Wang1, Daniel Leung, Fred Butterworth.
Abstract
A method using liquid chromatography electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) for the determination of trace levels of five macrolide antibiotics (spiramycin, tilmicosin, oleandomycin, erythromycin, and tylosin) in eggs is presented. Data acquisition under MS/MS was achieved by applying multiple reaction monitoring (MRM) of two or three fragment ion transitions to provide a high degree of sensitivity and specificity for both quantification and confirmation. Matrix-matched standard calibration curves were used to achieve the best accuracy of the method. A fully nested experimental design was used to study the measurement uncertainty arising from intermediate precision and trueness or proportional bias. The overall recoveries, that is, those determined by the nested experiments, of spiramycin, tilmicosin, oleandomycin, erythromycin, and tylosin at fortified levels of 60, 100, 200, and 300 microg/kg were 96.8, 98.2, 98.3, 98.8, and 95.4%, respectively. The LC/ESI-MS/MS method detection limits (S/N > or = 3:1) of five macrolides were <1.0 microg/kg.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15769104 DOI: 10.1021/jf048414p
Source DB: PubMed Journal: J Agric Food Chem ISSN: 0021-8561 Impact factor: 5.279