Inhomogeneity of ventilation leads to unpredictable errors in measured D(L)CO.

Bruce R Thompson1, G Kim Prisk, Philip Peyton, Robert J Pierce, Peter D Rochford.   

Abstract

We evaluated the effects of inhomogeneity of ventilation on single-breath (SB), rebreathing (RB) and open circuit (OC) D(L)CO using a mathematical model consisting of two alveolar compartments and a common dead space. Inhomogeneity in ventilation was studied by altering inspired volume, initial alveolar volume and compartment size independently. When distribution of inspired volume between alveolar compartments was inhomogeneous (9:1), D(L)CO was underestimated by 35% for SB, 25% for RB, and 16% for OC, and there was an underestimation in V(A) of 9%, 15% and 9%, respectively. With inhomogeneity in initial alveolar volume there was an overestimation in D(L)CO of 13%, 7% and 11% for SB, RB and OC techniques and an underestimation of V(A) of 7%, 12% and 9%. Finally inhomogeneity of compartment size led to an underestimation of D(L)CO of 18%, 35% and 36% with no change in measured V(A). These results suggest D(L)CO measurements are sensitive to inhomogeneity of ventilation, and importantly, all techniques were at times, significantly in error.

Entities:  

Mesh:

Year:  2005        PMID: 15766908     DOI: 10.1016/j.resp.2004.12.001

Source DB:  PubMed          Journal:  Respir Physiol Neurobiol        ISSN: 1569-9048            Impact factor:   1.931


  1 in total

1.  Endobronchial Valve Treatment in Emphysema Patients with a Very Low DLCO.

Authors:  Marlies van Dijk; Jorine E Hartman; Karin Klooster; Nick H T Ten Hacken; Huib A M Kerstjens; Dirk-Jan Slebos
Journal:  Respiration       Date:  2020-01-21       Impact factor: 3.580

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.