LC-MS/MS systematic toxicological analysis: comparison of MS/MS spectra obtained with different instruments and settings.

Robert Jansen1, Gérard Lachatre, Pierre Marquet.   

Abstract

OBJECTIVES: To compare the mass spectra obtained using a linear-ion-trap (LIT) tandem mass spectrometer (QTRAP) operated in the "enhanced product ion scan" (EPI) mode with those obtained in the classical triple-quadrupole product ion scan (PIS) mode run on the same as well as on two other instruments (TSQ-Quantum and Quattro-Micro). DESIGN AND METHODS: After tentative standardization of ion fragmentation and transmission in both polarities using a reference compound (glafenine) on the three instruments, eight test compounds detected in the positive mode and five in the negative mode were systematically infused in different ionization sources and spectral acquisition performed over approximately 5 s. The relative intensity of the ions present in the resulting spectra was quantitatively and statistically compared. Also, the intra-day and inter-day variabilities of these relative intensities, as well as the effect of increasing compound concentration, were studied using QTRAP operated in EPI mode.
RESULTS: The EPI and PIS modes operated on a single LIT MS/MS instrument resulted in significant differences in relative ion intensities in both polarities, and so did the other two instruments despite prior standardization with glafenine. Some fragments could be absent in certain spectra, but no unexpected or unique fragments showed up. Intra-day variability was smaller in the LIT EPI than in the regular PIS mode and in the positive than in the negative polarity. In EPI mode, both intra- and inter-day variabilities increased when the relative intensity decreased. The effect of increasing concentration on the relative intensity of major and minor ions was small but significant in both polarities. Finally, contamination and cleansing of the ionization source also had noticeable effects on MS/MS spectra, though the cause is unclear.
CONCLUSIONS: MS/MS spectra do not offer the expected inter-instrument reproducibility despite an attempt at standardizing the fragmentation conditions using a reference compound. However, although the inter-instrument differences in ion relative intensity were significant, the spectra obtained looked almost similar. This suggests that in library searching algorithms, higher weight should be assigned for the m/z ratios than for their relative intensity in the spectra.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15766737     DOI: 10.1016/j.clinbiochem.2004.11.003

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  7 in total

Review 1.  Mass spectrometry-based metabolomics.

Authors:  Katja Dettmer; Pavel A Aronov; Bruce D Hammock
Journal:  Mass Spectrom Rev       Date:  2007 Jan-Feb       Impact factor: 10.946

Review 2.  Modern instrumental methods in forensic toxicology.

Authors:  Michael L Smith; Shawn P Vorce; Justin M Holler; Eric Shimomura; Joe Magluilo; Aaron J Jacobs; Marilyn A Huestis
Journal:  J Anal Toxicol       Date:  2007-06       Impact factor: 3.367

3.  A method for monitoring and controlling reproducibility of intensity data in complex electrospray mass spectra: a thermometer ion-based strategy.

Authors:  Paolo Lecchi; Jinghua Zhao; Wesley S Wiggins; Tzong-Hao Chen; Ping F Yip; Brian C Mansfield; John M Peltier
Journal:  J Am Soc Mass Spectrom       Date:  2008-11-06       Impact factor: 3.109

4.  Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry.

Authors:  James J Pitt
Journal:  Clin Biochem Rev       Date:  2009-02

5.  Performance evaluation of three liquid chromatography mass spectrometry methods for broad spectrum drug screening.

Authors:  Kara L Lynch; Autumn R Breaud; Hilde Vandenberghe; Alan H B Wu; William Clarke
Journal:  Clin Chim Acta       Date:  2010-06-09       Impact factor: 3.786

6.  jTraML: an open source Java API for TraML, the PSI standard for sharing SRM transitions.

Authors:  Kenny Helsens; Mi-Youn Brusniak; Eric Deutsch; Robert L Moritz; Lennart Martens
Journal:  J Proteome Res       Date:  2011-10-13       Impact factor: 4.466

7.  Development of quantitation method for glycated aminophospholipids at the molecular species level in powdered milk and powdered buttermilk.

Authors:  Ai Kodate; Yurika Otoki; Naoki Shimizu; Junya Ito; Shunji Kato; Naoki Umetsu; Teruo Miyazawa; Kiyotaka Nakagawa
Journal:  Sci Rep       Date:  2018-06-07       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.