| Literature DB >> 15766196 |
Jin-wei Zhao1, Wei Cheng, Ya-hui Feng.
Abstract
An analytical method for determining wear metal elements directly by ICP-AES was proposed in this paper. The optimum condition was selected. The Conostan S-21 and Conostan 75 (oil-base standard samples) prepared by multi-element oil-base reserved standard samples were applied to make calibration curve. Oil samples were diluted 5 times by kerosene. The contents of wear metals in lubrication oil were determined by ICP-AES. The relative standard deviation (RSD) of all elements was less than 1%. In addition, the authors represented the analytical efficacy of Leeman labs multi-track profile ICP, and its rapid and accurate characters of simultaneous multi-element measurement.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15766196
Source DB: PubMed Journal: Guang Pu Xue Yu Guang Pu Fen Xi ISSN: 1000-0593 Impact factor: 0.589