Gary H Kramer1, Barry M Hauck, Kevin Capello.
Abstract
The performance characteristics of a commercially available whole body counting phantom have been examined experimentally and using Monte Carlo simulations. The counting efficiency as a function of photon energy (126 keV to 2,754 keV) obtained using this phantom has been compared with those obtained using a real or virtual BOMAB phantom in a scanning detector and a FastScan whole body counter. The results suggest that the commercially available whole body counting phantom is equivalent to a BOMAB phantom under certain circumstances but that under other measurement conditions it may be necessary to develop correction factors.Mesh:
Substances:
Year: 2005 PMID: 15761296 DOI: 10.1097/01.hp.0000152106.93202.05
Source DB: PubMed Journal: Health Phys ISSN: 0017-9078 Impact factor: 1.316