Literature DB >> 15759581

A clinical water-coated antenna applicator for MR-controlled deep-body hyperthermia: a comparison of calculated and measured 3-D temperature data sets.

Jacek Nadobny1, Waldemar Wlodarczyk, Lothar Westhoff, Johanna Gellermann, Roland Felix, Peter Wust.   

Abstract

A magnetic resonance (MR)-compatible three-dimensional (3-D) hyperthermia applicator was developed and evaluated in the magnetic resonance (MR) tomograph Siemens MAGNETOM Symphony 1.5 T. Radiating elements of this applicator are 12 so-called water coated antenna (WACOA) modules, which are designed as specially shaped and adjustable dipole structures in hermetically closed cassettes that are filled by deionized water. The WACOA modules are arranged in the applicator frame in two transversal antenna subarrays, six antennas per subarray. As a standard load for the applicator an inhomogeneous phantom was fabricated. Details of applicator's realization are presented and a 3-D comparison of calculated and measured temperature data sets is made. A fair agreement is achieved that demonstrates the numerically supported applicator's ability of phase-defined 3-D pattern steering. Further refinement of numerical models and measuring methods is necessary. The applicator's design and the E-field calculations were performed using the finite-difference time-domain (FDTD) method. The calculation and optimization of temperature patterns was obtained using the finite element method (FEM). For MR temperature measurements the proton resonance frequency (PRF) method was used.

Entities:  

Mesh:

Year:  2005        PMID: 15759581     DOI: 10.1109/TBME.2004.843291

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Mathematical formulation and analysis of the nonlinear system reconstruction of the online image-guided adaptive control of hyperthermia.

Authors:  Kung-Shan Cheng; Mark W Dewhirst; Paul F Stauffer; Shiva Das
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Fast PRF-based MR thermometry using double-echo EPI: in vivo comparison in a clinical hyperthermia setting.

Authors:  Tetiana Dadakova; Johanna Gellermann; Otilia Voigt; Jan Gerrit Korvink; John Matthew Pavlina; Jürgen Hennig; Michael Bock
Journal:  MAGMA       Date:  2014-11-08       Impact factor: 2.310

3.  Feasibility of A-mode ultrasound attenuation as a monitoring method of local hyperthermia treatment.

Authors:  Noraida Abd Manaf; Maizatul Nadwa Che Aziz; Dzulfadhli Saffuan Ridzuan; Maheza Irna Mohamad Salim; Asnida Abd Wahab; Khin Wee Lai; Yan Chai Hum
Journal:  Med Biol Eng Comput       Date:  2016-04-02       Impact factor: 2.602

  3 in total