This paper describes the measurement of scattered radiation from a computed tomography (CT) scanner in a clinical situation and compares the results with those obtained from a CT performance phantom and with data obtained from CT manufacturers. The results are presented as iso-dose contours. There are significant differences between the data obtained and that supplied by manufacturers, both in the shape of the iso-dose contours and in the nominal values. The observed scatter in a clinical situation (for an abdominal scan) varied between 3% and 430% of the manufacturers' stated values, with a marked reduction in scatter noted at the head and feet of the patient. These differences appear to be due to the fact that manufacturers use CT phantoms to obtain scatter data and these phantoms do not provide the same scatter absorption geometry as patients. CT scatter was observed to increase as scan field size and slice thickness increased, whilst there was little change in scatter with changes in gantry tilt and table slew. Using the iso-dose contours, the orientation of the CT scanner can be optimised with regard to the location and shielding requirements of doors and windows. Additionally, the positioning of staff who must remain in the room during scanning can be optimised to minimise their exposure. The data presented here allows for realistic radiation protection assessments to be made.Entities: