Lizhong Liu1, Satish C Prasad, Daniel A Bassano, Joel Heavern, Bonnie Keshler, Seung S Hahn.
Abstract
Misplacement of dwell positions is a potential source of misadministration in high dose rate brachytherapy. In this work we present a dwell position verification method using fluoroscopic images. A mobile C-arm fluoroscopic machine is used to take a snapshot of the treatment machine's check cable as it reaches the most distal dwell position. This fluoroscopic image is displayed side-by-side with a treatment planning image on a dual monitor relay station at the HDR treatment console. Any discrepancy between the check cable's position on the verification image and the intended dwell position on the planning image can be identified immediately, thus avoiding the possibility of treating the wrong target volume.Entities:
Mesh:
Year: 2004 PMID: 15753927 PMCID: PMC5723442 DOI: 10.1120/jacmp.v5i1.1944
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Schematic diagram showing the C‐arm and its dual monitor workstation in the HDR treatment room. The relay station is shown outside the HDR treatment room near the console. One of the monitors of the workstation as well as the relay station is used to display the treatment planning image. The other monitor is used for verification image. A C‐arm remote control pendant is available at the relay station for taking fluoroscopic images of the HDR check cable.
Figure 2Application of our verification method to a two‐catheter implant in a phantom. (a) Treatment planning image taken in the AP orientation with the first and the last dwell positions of the source marked using arrows. The distance scale is indicated by the dummy seeds, spaced 1 cm apart, and is marked in the figure. (b) Verification image of the check cable at the first dwell position compared to the first dwell position in Fig. 2(a).
Figure 3Example of our verification method applied to a 16‐catheter interstitial gynecological implant. (a) Reference image taken in the AP orientation with the first and last dwell positions for the first catheter marked. The distance scale is indicated by the dummy seeds spaced at 1 cm apart and is marked in the figure. (b) Verification image of the check cable at the first dwell position of the first catheter.