Literature DB >> 15751676

ACR II: improved absolute cryogenic radiometer for low background infrared calibrations.

Adriaan C Carter1, Steven R Lorentz, Timothy M Jung, Raju U Datla.   

Abstract

A second-generation absolute cryogenic radiometer (ACR II) was developed for use at the Low Background Infrared calibration facility at the National Institute of Standards and Technology. The need for spectral calibrations of very sensitive [D* = 10(14) cm (Hz)1/2W(-1)] infrared detectors necessitated the use of a cryogenic infrared monochromator and a more sensitive radiometer. The improved low-power performance of the ACR II compared with the older absolute cryogenic radiometer (ACR) has also made it useful as the primary standard for the calibration of cryogenic blackbody sources that are used as low-power infrared sources. The responsivity of the new radiometer's receiver is 210 K/mW with a type A (random component) standard uncertainty of at most 7 pW when making power measurements of less than 10 nW. The original ACR has a responsivity of 29 K/mW and has a type A standard uncertainty of approximately 100 pW when making a similar low-noise-power measurement. Other properties of the radiometers are also described and compared.

Year:  2005        PMID: 15751676     DOI: 10.1364/ao.44.000871

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  Intercomparison of the LBIR Absolute Cryogenic Radiometers to the NIST Optical Power Measurement Standard.

Authors:  James A Fedchak; Adriaan C Carter; Raju Datla
Journal:  J Res Natl Inst Stand Technol       Date:  2006-08-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.