An in vitro evaluation of ZiReal abutments with hexagonal connection: in original state and following abutment preparation.

Paolo Vigolo1, Fulvio Fonzi, Zeina Majzoub, Giampiero Cordioli.   

Abstract

PURPOSE: Laboratory processing of implant-supported prostheses may alter the surface of the abutment in contact with the implant head, with potential repercussions for the interface fit. The purpose of this study was to assess changes at the implant interface of high-strength zirconia ceramic esthetic abutments with a hexagonal connection (ZiReal; 3i/Iimplant Innovations, Palm Beach Gardens, FL) following abutment preparation for single-tooth restorations.
MATERIALS AND METHODS: The depth (d) and width (w) of the titanium hexagonal portion of the abutment, the apical diameter of the abutment (D), and the rotational freedom (R) of the abutment were assessed for 20 ZiReal abutments prior to preparation (time 0) and following abutment preparation (time 1) to detect any eventual change of fit of the abutment on the top of the implant hexagon.
RESULTS: No significant differences relative to any study parameter (d, w, D, and R) were observed between time 0 and time 1 (P = .9542). DISCUSSION AND
CONCLUSIONS: The hexagonal misfit of the titanium machined ZiReal abutment on the implant hexagon may be implicated in screw joint loosening. The results of this report suggest that if all laboratory steps are carefully observed, changes at the implant/ZiReal abutment do not occur. The maintenance of the original features of the ZiReal abutment may reduce the risk of screw loosening.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15747681

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  2 in total

1.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

2.  Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Authors:  Mitra Zirak; Mahroo Vojdani; Amir Ali Reza Khaledi; Mitra Farzin
Journal:  J Dent (Shiraz)       Date:  2019-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.