Literature DB >> 15747209

Deep brain stimulation in intraoperative MRI environment - comparison of imaging techniques and electrode fixation methods.

M W Y Lee1, A A F De Salles, L Frighetto, R Torres, E Behnke, J M Bronstein.   

Abstract

We performed 118 consecutive DBS cases from November 1999 to June 2002. Intraoperatively there were 10 cases studied with fluoroscopy, 73 with 0.2 Tesla (T) MRI and 35 with 1.5 T MRI. Ten electrodes were secured by Medtronic caps, 25 by methyl methacrylate with titanium miniplates, and 82 by Navigus caps. The 3-dimensional displacement between the planned target and actual electrode position (3DD) was determined by fusing the postoperative MRI with the preoperative imaging. The 3DD for using Medtronic caps, methyl methacrylate with miniplates, and Navigus caps were 4.80 +/- 3.16, 2.64 +/- 1.26 and 2.23 +/- 1.15 mm (mean +/- SD), respectively. Navigus caps had statistically significant accuracy (P = 0.03) in holding the electrode when compared with Medtronic caps, and it facilitated electrode revision. The fixation devices significantly affect the final vertical position of the electrode. The 3DD for fluoroscopy, 0.2 T and 1.5 T MRI cases were 4.80 +/- 3.16, 2.31 +/- 1.21 and 2.34 +/- 1.14 mm (mean +/- SD), respectively. No statistically significant difference (P = 0.91) in 3DD was demonstrated between 0.2 T and 1.5 T MRI cases. The presence of intraoperative 1.5 T MRI allowed near real-time electrode position confirmation and early detection of hemorrhagic complications. Satisfactory microelectrode recording was feasible in low-field 0.2 T and high-field 1.5 T MRI environments. Further studies on performing DBS in real-time intraoperative MRI are warranted.

Entities:  

Mesh:

Year:  2005        PMID: 15747209     DOI: 10.1055/s-2004-830169

Source DB:  PubMed          Journal:  Minim Invasive Neurosurg        ISSN: 0946-7211


  7 in total

1.  Deep brain stimulation for movement and other neurologic disorders.

Authors:  Mahlon DeLong; Thomas Wichmann
Journal:  Ann N Y Acad Sci       Date:  2012-07-23       Impact factor: 5.691

2.  Lateralization of the Subthalamic Nucleus with Age in Parkinson's Disease.

Authors:  Julio L B Pereira; Sydney Furie B A; Justin Sharim; Daniel Yazdi; Antonio A F DeSalles; Nader Pouratian
Journal:  Basal Ganglia       Date:  2016-04-01

3.  Diffusion tensor imaging (DTI) and colored fractional anisotropy (FA) mapping of the subthalamic nucleus (STN) and the globus pallidus interna (GPi).

Authors:  Mark Sedrak; Alessandra Gorgulho; Ausaf Bari; Eric Behnke; Andrew Frew; Inga Gevorkyan; Nader Pouratian; Antonio DeSalles
Journal:  Acta Neurochir (Wien)       Date:  2010-10-03       Impact factor: 2.216

4.  A role of diffusion tensor imaging in movement disorder surgery.

Authors:  Garni Barkhoudarian; Tony Klochkov; Mark Sedrak; Andrew Frew; Alessandra Gorgulho; Eric Behnke; Antonio De Salles
Journal:  Acta Neurochir (Wien)       Date:  2010-07-21       Impact factor: 2.216

5.  Subthalamic nucleus deep brain stimulator placement using high-field interventional magnetic resonance imaging and a skull-mounted aiming device: technique and application accuracy.

Authors:  Philip A Starr; Alastair J Martin; Jill L Ostrem; Pekka Talke; Nadja Levesque; Paul S Larson
Journal:  J Neurosurg       Date:  2010-03       Impact factor: 5.115

6.  The pros and cons of intraoperative CT scan in evaluation of deep brain stimulation lead implantation: A retrospective study.

Authors:  Domenico Servello; Edvin Zekaj; Christian Saleh; Claudio Pacchetti; Mauro Porta
Journal:  Surg Neurol Int       Date:  2016-08-02

7.  Dual floor burr hole technique in deep brain stimulation: A retrospective study on 209 patients.

Authors:  Domenico Servello; Christian Saleh; Alberto R Bona; Marina Minichiello; Edvin Zekaj
Journal:  Surg Neurol Int       Date:  2017-11-20
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.