Doppler echocardiography of 119 normal-functioning St Jude Medical mitral valve prostheses: a comprehensive assessment including time-velocity integral ratio and prosthesis performance index.

Joseph F Malouf1, Manfredi Ballo, Heidi M Connolly, David O Hodge, Regina M Herges, Charles J Mullany, Fletcher A Miller.   

Abstract

The purpose of this study was to provide, in a large number of patients, comprehensive Doppler echocardiographic assessment of normal St Jude Medical mitral valve prosthesis function using Doppler-derived hemodynamic variables, including the mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio and prosthesis performance index. The pressure half-time was less than 130 milliseconds in all patients, and all but one patient had either a peak early mitral diastolic velocity of 2 m/s or less or a mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio of less than 2.2. There was a significant (P < .001) negative correlation between the prosthesis performance index and prosthesis size. This negative correlation suggests that there is more efficient use of the in vitro geometric orifice area with smaller prostheses.

Entities:  

Mesh:

Year:  2005        PMID: 15746715     DOI: 10.1016/j.echo.2004.11.011

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  1 in total

1.  Echocardiographically derived effective valve opening area in mitral prostheses: a comparative analysis of various calculations using continuity equation and pressure half time method.

Authors:  Nikola Bogunovic; Dieter Horstkotte; Lothar Faber; Lukas Bogunovic; Frank van Buuren
Journal:  Heart Vessels       Date:  2015-06-07       Impact factor: 2.037

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.