Literature DB >> 15743574

[Evaluation of the Robobact system for processing stool cultures: another step towards automation in the microbiology laboratory].

Mercedes Alonso-Sanz1, Laura Molina.   

Abstract

OBJECTIVE: The objective of this study was to evaluate the Robobact system (DIESSE Diagnostica Senese S.p.A., Italy) for processing stool specimens and to compare it with conventional methodology.
METHOD: A total of 240 stool specimens from outpatients in Madrid (Spain) were studied. The samples were processed simultaneously with both the Robobact system and a conventional method.
RESULTS: Overall, 13 isolates of Campylobacter spp., 12 of Salmonella spp., 4 of Yersinia enterocolitica and 1 of Shigella spp. were obtained. The Robobact method failed to identify 1 of the 13 Campylobacter spp. isolates and 2 of the 12 Salmonella spp. isolates. The single Shigella spp. isolate was detected only by the Robobact system. All the Robobact results were available at 24 hours.
CONCLUSIONS: Robobact is a fast, useful tool for microbiological processing of stool specimens.

Entities:  

Mesh:

Year:  2005        PMID: 15743574     DOI: 10.1157/13071606

Source DB:  PubMed          Journal:  Enferm Infecc Microbiol Clin        ISSN: 0213-005X            Impact factor:   1.731


  1 in total

1.  Evaluation of two automated systems for detection of bacteriuria.

Authors:  José Gutiérrez; Antonio Sorlózano; María José Soto; Gonzalo Piédrola
Journal:  J Clin Lab Anal       Date:  2006       Impact factor: 2.352

  1 in total